G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-02944
- Event Type
- Malfunction
- Date Received
- April 16, 2015
- Date of Event
- March 20, 2015
- Report Date
- March 20, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER WAS NOT RETURNED TO DEXCOM. THE PEDIATRIC RECEIVER ((B)(4) LOT NUMBER 5191848), BEING USED WITH THE COMPLAINT TRANSMITTER, WAS RETURNED ON (B)(6) 2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT FO AN INTERMITTENT OUT OF RANGE SIGNAL. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION. HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. THE DATA WAS REVIEWED ON 04/09/2015 AND CONFIRMED THE REPORTED EVENT OF INTERMITTENT OUT OF RANGE.
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT INTERMITTENT OUT OF RANGE SIGNAL ON (B)(6) 2015. THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255936 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-05 | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 7 YR |