SIMULATOR HP
MDR report key: 469446
·
Received June 27, 2003
Report
- Report Number
- 1121753-2003-00002
- Event Type
- Other
- Date Received
- June 27, 2003
- Date of Event
- April 30, 2003
- Report Date
- June 25, 2003
- Manufacturer
- NUCLETRON BV
- Product Code
- KPQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
DURING ACQUISITION OF SIMULATOR CT IMAGES THE GANTRY STOPPED. UNDER GRAVITY THE FILM CASSETTE HAD SLIPPED AND RELEASED THE LOCKING HANDLES OF THE CT AFTER-SPLIT. SINCE THE AFTER-SLIT WAS NOT IN THE CORRECT POSITION, THE GANTRY WOULD NOT ROTATE. TO CORRECT THE PROBLEM, THE TECHNOLOGIST TRIED TO ADJUST THE FILM CASSETTE IN POSITION. IN DOING SO, THE CT AFTER-SLIT FELL AND THE EDGE OF THE AFTER-SLIT HIT THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIMULATOR HP | RADIATION THERAPY SIMULATOR WITH CT EXTENSION | KPQ | NUCLETRON BV | HP MARK 1/CTX | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Other |