FDA Adverse Event Malfunction Summary report: N

CUAS EXCEL 23KHZ CEM NOSECONE

MDR report key: 4692955 · Received April 13, 2015

Report

Report Number
3006697299-2015-00048
Event Type
Malfunction
Date Received
April 13, 2015
Report Date
March 19, 2015
Manufacturer
INTEGRA NEUROSCIENCES PR
Product Code
LFL
PMA / PMN Number
K141674
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVAL. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO.

Additional Manufacturer Narrative · 1

INTEGRA COMPLETED ITS INTERNAL INVESTIGATION 8/31/2015. THE INVESTIGATION INCLUDED: METHOD: -DHR REVIEW, -COMPLAINT TREND, -VISUAL INSPECTION. RESULTS: MANUFACTURING DATES: LOT: 1140146 - 2/8/2014, LOT: 1141946 - 6/29/2014. THE DHRS OF THE FG LOTS # 1140146 AND 1141946 WERE REVIEWED AS PART OF THIS INVESTIGATION. THE FG LOTS WERE RELEASED FOR DISTRIBUTION PURPOSE IN COMPLIANCE WITH THE PRODUCT SPECIFICATIONS AND QUALITY SYSTEM REQUIREMENTS. A TWO YEAR LOOK BACK FOR THIS REPORTED FAILURE AND OR RELATED TO "DIATHERMY CONTINUED AFTER BUTTON RELEASED" FOR THIS PRODUCT ID SHOWS THAT 26 COMPLAINTS WERE RECEIVED INCLUDING THIS CASE. A CAPA WAS ISSUED TO FURTHER INVESTIGATE THIS REPORTED FAILURE. CONCLUSION: REPORTED FAILURE COULD NOT BE CONFIRMED OR REPRODUCED. HOWEVER THE BUTTON DID NOT ACTIVATE CEM WHEN PUSHED, WHICH DOES INDICATE A POSSIBLE ISSUE WITH THE SWITCH. LEAKAGE CURRENT WAS BELOW THE ACCEPTABLE LEVEL OF 2000MO (READ 1.5MO), WHICH MAY BE DUE TO A BREAKDOWN IN THE INSULATING PROPERTY OF THE SWITCH/ACTIVATION MECHANISM OF THE NOSECONE, AS OPPOSED TO ANOTHER PART OF THE CEM CIRCUIT. ADDITIONALLY, THE SWITCH CIRCUIT WAS READING GREATER THAN 2MO (READ 11.4MO), WHICH INDICATES AN OPEN CIRCUIT CONDITION IN THE CEM SWITCH. BUILDUP OF WHITE MATERIAL WAS NOTICED IN THE SWITCH, WHICH DOES NOT ALLOW SWITCH TO PRESS DOWN PROPERLY AND MAY BE INDICATIVE OF MATERIAL MIGRATION.

Description of Event or Problem · 1

THE FIRST REPORT OF TWO INVOLVING THE SAME PRODUCT C6623 CUSA EXCEL CEM 23KHZ NOSECONE AND THE SAME PT. IT WAS REPORTED THAT THE DIATHERMY CONTINUED AFTER THE BUTTON WAS RELEASED DURING AN OPEN HEPATECTOMY-- SO THEREFORE NOTABLE STICKING. THERE WERE 2 NOSECONES WITH THE SAME ISSUE DURING THE 1 PROCEDURE- LOT# 1140146 X 1; LOT# 1141946 X 1. DELAY TO SURGERY APPROX 20 MINUTES WHILE THEY CHANGED OVER NOSECONES. NO ADVERSE PT OUTCOMES WERE REPORTED. THE SALES REP STATED THAT THIS SURGEON/HOSP ARE STILL AWARE OF THE ISSUE ABOUT CHANGING OVER THE NOSECONE AFTER 30 MINUTES AS PER IFU INSTRUCTIONS. SALES REP NOTED THAT IT WAS A DIFFICULT CASE THAT WAS PROLONGED AND DIATHERMY WAS ON FOR MORE THAN 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
244145 CUAS EXCEL 23KHZ CEM NOSECONE ULTRASONIC SURGICAL PRODUCTS LFL INTEGRA NEUROSCIENCES PR 1141946

Patients

Seq Age Sex Outcome Treatment
1 60 YR