FDA Adverse Event Malfunction Summary report: N

ONE TOUCH ULTRA

MDR report key: 468982 · Received June 23, 2003

Report

Report Number
2939301-2003-04051
Event Type
Malfunction
Date Received
June 23, 2003
Report Date
June 16, 2003
Manufacturer
LIFESCAN, INC.
Product Code
CFR
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

PT CALLED LFS IN 2003 ALLEGING THAT THEIR METER WAS READING INACCURATELY HIGH. MEDICAL AFFAIRS SPECIALIST SPOKE WITH PT THE NEXT DAY TO OBTAIN ADDITIONAL INFO. PT EXPLAINED THAT THEY OBTAINED A "210 MG/DL" AT 6:30 AM. PT PROCEEDED TO TAKE 4 UNITS OF HUMALOG AND 18 UNITS OF NPH. THEY WERE NOT EXPERIENCING ANY HYPERGLYCEMIC SYMPTOMS AT THE TIME. DUE TO THE "210 MG/DL", PT DECIDED TO EAT FEWER CARBOHYDRATES THAN NORMAL. APPROXIMATELY FIVE AND HALF HOURS LATER, PT STARTED FEELING SLEEPY, SWEATY, DISORIENTED, WITH SHAKY LEGS, SEEING SPOTS AND EVENTUALLY BLACKED OUT. PT EXPLAINED THAT SPOUSE FED THEM A TUBE (.68 OZ) OF CAKEMATE DECORATING GEL. PT DID NOT RECEIVE ANY MEDICAL CARE FROM A HEALTHCARE PROFESSIONAL. THE NEXT BLOOD GLUCOSE TEST WAS PERFORMED AT 5:45 PM. A RESULT OF "277 MG/DL" WAS OBTAINED ON THE REPORTED METER. PT REPORTED THAT THEIR TEST STRIPS HAD BEEN PEELING UPON INSERTION INTO THE METER. PT STATED THAT THEY HAD BEEN EXPERIENCING THE PEELING STRIPS WITH MANY DIFFERENT VIALS FOR AN UNK LENGTH OF TIME. PT REPORTED THAT THE TEST STRIP PEELED WHEN PT OBTAINED THE "210 MG/DL" READING THE DAY OF THE INCIDENT. THE CUSTOMER SERVICE REP WALKED PT THROUGH A CONTROL SOLUTION TEST, WHICH FELL OUTSIDE THE ACCEPTED RANGE (193/109-148). THERE IS NO EVIDENCE TO SUGGEST THAT THE PT SUFFERED THE ADVERSE EVENT DUE TO THE METER. PT OBTAINED A READING AND TOOK THE PRESCRIBED AMOUNT OF INSULIN (NO EXTRA UNITS), HOWEVER, PT DECIDED ON THEIR OWN TO EAT FEWER CARBOHYDRATES DUE TO THE "210 MG/DL" READING. FIVE AND HALF-HOURS LATER THEY EXPERIENCED WHAT SEEMED TO BE HYPOGLYCEMIC SYMPTOMS, HOWEVER, THEY DID NOT PERFORM A TEST UNTIL 5 HOURS AFTER THEY EXPERIENCED SYMPTOMS. THE CUSTOMER SERVICE REPLACED THE METER AND CONTROL SOLUTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING KIT/SYSTEM CFR LIFESCAN, INC. NA 1021733

Patients

Seq Age Sex Outcome Treatment
1 74 YR