FX OXY W/HR & ART FILTER RESERVOIR
Report
- Report Number
- 9681834-2015-00061
- Event Type
- Injury
- Date Received
- April 8, 2015
- Date of Event
- March 23, 2015
- Report Date
- March 25, 2015
- Manufacturer
- TERUMO CORP, ASHITAKA
- Product Code
- DTZ
- PMA / PMN Number
- K112229
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
TERUMO HAS NOT RECEIVED THE ACTUAL DEVICE FOR EVALUATION; THEREFORE, THE INVESTIGATION HAS YET TO BE COMPLETED. TERUMO PLANS ON SUBMITTING A FOLLOW-UP REPORT WHEN THE INVESTIGATION IS COMPLETE AND MORE INFORMATION BECOMES AVAILABLE. FOR THIS REASON, TERUMO REFERENCES EVALUATION CONCLUSION. (B)(4).
THE USER FACILITY REPORTED TO TERUMO CARDIOVASCULAR SYSTEMS CORP THAT UPON INITIATING CARDIOPULMONARY BYPASS, THEY EXPERIENCED FLOW ISSUES WHILE USING THE FX15 OXYGENATOR. PRIMING OF THE CIRCUIT HAD NO ISSUES. THE PATIENT HAD A BSA OF 1.98 AND WAS EMERGENTLY PLACED ON CPB AFTER BECOMING HEMODYNAMICALLY UNSTABLE. AS CPB WAS INITIATED, THE TARGET BLOOD FLOW RATE WAS 4.8 L/MIN, WHILE THE MAXIMUM FLOW RATE PRODUCED WAS ONLY 3.6 L/MIN, EVEN THOUGH THE RPM HAD BEEN GRADUALLY RAISED TO 3600 RPM. DURING THE NEXT 5 MINUTES, THE FLOW RATE HAD DECREASED TO ABOUT 2.5 - 2.75 L/MIN. THE CIRCUIT LINE PRESSURE WAS MEASURED JUST AFTER THE OXYGENATOR AT NORMAL VALUES OF 180-200MMHG. VARIOUS ADJUSTMENTS WERE MADE WITH NO AFFECT SO THE UNIT WS CHANGED OUT RESULTING IN THE PATIENT LOSING 200-300CC OF BLOOD THAT REMAINED IN THE ORIGINAL CIRCUIT. THE PROCEDURE WAS DELAYED APPROXIMATELY 40 MINS DUE TO THE ISSUE AND PRODUCT CHANGEOUT. THERE WERE NO ISSUES THROUGHOUT THE REMAINDER OF THE PROCESS. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO PATIENT HARM WAS OBSERVED. SURGERY WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 232644 | FX OXY W/HR & ART FILTER RESERVOIR | BLOOD GAS OXYGENATOR | DTZ | TERUMO CORP, ASHITAKA | 1CX*FX15E | 140715 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |