FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4675459 · Received April 9, 2015

Report

Report Number
3004753838-2015-13054
Event Type
Malfunction
Date Received
April 9, 2015
Date of Event
March 13, 2015
Report Date
March 13, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT SENSOR DEVICE WAS NOT RETURNED TO DEXCOM. THE TRANSMITTER DEVICE (9438-05/68U4D), USED WITH THE COMPLAINT SENSOR DEVICE, WAS RETURNED ON (B)(4) 2015. THE RETURNED COMPLAINT TRANSMITTER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. THE REPORTED EVENT OF INACCURATE BLOOD GLUCOSE VALUES COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT CONTINUOUS GLUCOSE MONITORING (CGM) INACCURACIES COMPARED TO BLOOD GLUCOSE METER ON (B)(6) 2015. AT THE TIME OF CONTACT, THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
234210 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9500-27 5197463

Patients

Seq Age Sex Outcome Treatment
1 4 YR