FDA Adverse Event Death Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4672859 · Received April 7, 2015

Report

Report Number
3008642652-2015-01781
Event Type
Death
Date Received
April 7, 2015
Date of Event
March 4, 2015
Report Date
March 31, 2015
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MONITOR SN (B)(4) AND BELT SN 59078409 HAVE NOT BEEN RETURNED FOR EVALUATION. DEVICE EVALUATION WAS ACCOMPLISHED THROUGH A REVIEW OF THE PATIENT'S DOWNLOADED DATA FILE. REVIEW OF THE DATA DOES NOT INDICATE ANY DEVICE MALFUNCTION RELATED TO THE EVENT. H4: MONITOR (B)(4): 04/2012. ELECTRODE BELT (B)(4): 04/2014.

Description of Event or Problem · 1

A DISTRIBUTOR CONTACTED ZOLL ON (B)(6) 2015 TO REPORT THAT A PATIENT PASSED AWAY ON (B)(6) 2015 AT 5:17 PM. PER THE DISTRIBUTOR, THE PATIENT WAS WITH HIS WIFE ON HIS WAY HOME FROM (B)(6) AND WAS NOT FEELING WELL. HIS WIFE TOOK HIM TO A HOSPITAL IN AL WHERE HE WAS TAKEN TO THE ICU AND LATER DIED. THE PATIENT'S WIFE REPORTED HE WAS TREATED WHILE IN THE ER. AFTER THAT SHE WAS TOLD HE WAS "BRAIN DEAD AND HIS HEART THEN STOPPED." INVESTIGATION INTO THE EVENT REVEALED THAT THE PATIENT EXPERIENCED AN INAPPROPRIATE DEFIBRILLATION EVENT ON (B)(6) 2015. THE TREATMENT WAS DELIVERED AT 01:28:42 AM. THE RHYTHM AT THE TIME OF THE TREATMENT WAS MOTION ARTIFACT, WHICH CONTRIBUTED TO THE FALSE DETECTION. THE POST SHOCK RHYTHM WAS A SINUS RHYTHM AT 90 BPM WITH MOTION ARTIFACT. A NON-TREATABLE RHYTHM WAS DETECTED AT 01:29:05 AM AND THE DEVICE WAS SHUT DOWN AT 02:04:11 AM. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE ENTIRE EVENT. THE DEVICE WAS NOT ACTIVE AT THE REPORTED TIME OF PASSING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
228264 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR Death