FDA Adverse Event Malfunction Summary report: N

HIP END EFFECTOR- VARIABLE ANGLE

MDR report key: 4671987 · Received April 9, 2015

Report

Report Number
3005985723-2015-00050
Event Type
Malfunction
Date Received
April 9, 2015
Date of Event
March 11, 2015
Report Date
March 11, 2015
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K121064
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP#1 SUBMITTED TO CORRECT, BASED ON THE RESULTS OF INVESTIGATION. AN EVENT REGARDING A LOCKING MECHANISM FAILURE WITH THE PARALLEL HIP END EFFECTOR WAS CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE PARALLEL HIP END EFFECTOR EVENT WAS CONFIRMED. DEVICE HISTORY REVIEW: BASED ON THE DEVICE HISTORY REVIEW THERE WERE NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING LOCKING MECHANISM FAILURE OF CATALOG: 206967, SERIAL NUMBER: (B)(4), LOT #: 19060212. RMA #: (B)(4). THERE HAVE BEEN NO OTHER EVENTS FOR THE REFERENCED SERIAL NUMBER OR LOT NUMBER. CONCLUSIONS: THE PARALLEL HIP END EFFECTOR (P/N 206967, L/N 190179) WAS EVALUATED VISUALLY AND FUNCTIONALLY UPON RECEIPT AND THE REPORTED EVENT WAS CONFIRMED. THE PARALLEL HIP END EFFECTOR SHOWED A FAILED LOCKING MECHANISM KNOB WAS BROKEN, WHICH HAS MADE ACTUATING THE LOCKING MECHANISM DIFFICULT.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED AT MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE FILED WHEN ADDITIONAL INFORMATION IS OBTAINED.

Description of Event or Problem · 1

THE SURGEON WAS PERFORMING A MAKOPLASTY TOTAL HIP ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS PST HIP IMPLANTS. AFTER IMPACTING THE CUP, THE END EFFECTOR WOULD NOT DISENGAGE FROM THE IMPACTOR SHAFT. IT WAS NOTED THAT THE BACK OF THE HANDLE BROKE. THIS DID NOT AFFECT THE CASE AND IT WAS COMPLETED SUCCESSFULLY.

Description of Event or Problem · 1

THE SURGEON WAS PERFORMING A MAKOPLASTY TOTAL HIP ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS PST HIP IMPLANTS. AFTER IMPACTING THE CUP, THE END EFFECTOR WOULD NOT DISENGAGE FROM THE IMPACTOR SHAFT. IT WAS NOTED THAT THE BACK OF THE HANDLE BROKE. THIS DID NOT AFFECT THE CASE AND IT WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235080 HIP END EFFECTOR- VARIABLE ANGLE STEREOTAXIC DEVICE, ACCESSORY OLO MAKO SURGICAL CORP. 9060212

Patients

Seq Age Sex Outcome Treatment
1 Other