HIP END EFFECTOR- VARIABLE ANGLE
Report
- Report Number
- 3005985723-2015-00050
- Event Type
- Malfunction
- Date Received
- April 9, 2015
- Date of Event
- March 11, 2015
- Report Date
- March 11, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K121064
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP#1 SUBMITTED TO CORRECT, BASED ON THE RESULTS OF INVESTIGATION. AN EVENT REGARDING A LOCKING MECHANISM FAILURE WITH THE PARALLEL HIP END EFFECTOR WAS CONFIRMED. METHOD AND RESULTS: DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE PARALLEL HIP END EFFECTOR EVENT WAS CONFIRMED. DEVICE HISTORY REVIEW: BASED ON THE DEVICE HISTORY REVIEW THERE WERE NO REPORTED DISCREPANCIES. COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING LOCKING MECHANISM FAILURE OF CATALOG: 206967, SERIAL NUMBER: (B)(4), LOT #: 19060212. RMA #: (B)(4). THERE HAVE BEEN NO OTHER EVENTS FOR THE REFERENCED SERIAL NUMBER OR LOT NUMBER. CONCLUSIONS: THE PARALLEL HIP END EFFECTOR (P/N 206967, L/N 190179) WAS EVALUATED VISUALLY AND FUNCTIONALLY UPON RECEIPT AND THE REPORTED EVENT WAS CONFIRMED. THE PARALLEL HIP END EFFECTOR SHOWED A FAILED LOCKING MECHANISM KNOB WAS BROKEN, WHICH HAS MADE ACTUATING THE LOCKING MECHANISM DIFFICULT.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED AT MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE FILED WHEN ADDITIONAL INFORMATION IS OBTAINED.
THE SURGEON WAS PERFORMING A MAKOPLASTY TOTAL HIP ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS PST HIP IMPLANTS. AFTER IMPACTING THE CUP, THE END EFFECTOR WOULD NOT DISENGAGE FROM THE IMPACTOR SHAFT. IT WAS NOTED THAT THE BACK OF THE HANDLE BROKE. THIS DID NOT AFFECT THE CASE AND IT WAS COMPLETED SUCCESSFULLY.
THE SURGEON WAS PERFORMING A MAKOPLASTY TOTAL HIP ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS PST HIP IMPLANTS. AFTER IMPACTING THE CUP, THE END EFFECTOR WOULD NOT DISENGAGE FROM THE IMPACTOR SHAFT. IT WAS NOTED THAT THE BACK OF THE HANDLE BROKE. THIS DID NOT AFFECT THE CASE AND IT WAS COMPLETED SUCCESSFULLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235080 | HIP END EFFECTOR- VARIABLE ANGLE | STEREOTAXIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. | 9060212 |