FDA Adverse Event Injury Summary report: N

TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM

MDR report key: 4668673 · Received April 8, 2015

Report

Report Number
3005334138-2015-00027
Event Type
Injury
Date Received
April 8, 2015
Date of Event
March 12, 2015
Report Date
March 12, 2015
Manufacturer
ST. JUDE MEDICAL
Product Code
OAE
PMA / PMN Number
P130026
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. REVIEW OF THE LOG FILES REVEALED THE DEVICE FUNCTIONED AS INTENDED AND THERE WERE NO CONTRIBUTING ANOMALIES. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THE PROCESS WAS PERFORMED AND COMPLETED IN ACCORDANCE WITH SJM SPECIFICATIONS AND PROCEDURES. THE CAUSE OF THE REPORTED PERICARDIAL EFFUSION WAS UNABLE TO BE CONFIRMED AND REMAINS UNKNOWN. PER THE IFU, CARDIAC PERFORATION IS A KNOWN RISK WITH THE USE OF THIS DEVICE.

Description of Event or Problem · 1

DURING AN RE-DO ATRIAL FIBRILLATION ABLATION PROCEDURE, A PERICARDIAL EFFUSION OCCURRED. AFTER OBTAINING ACCESS TO THE LEFT ATRIUM THROUGH A PFO, A TACTICATH QUARTZ ABLATION CATHETER WAS USED TO ABLATE COMPLEX FRACTIONATED ELECTROGRAMS (CFES). DURING INITIAL ABLATION, THE PATIENT WENT INTO A TACHYCARDIA. CFE ABLATIONS WERE COMPLETED AND THE TACTICATH WAS WITHDRAWN BACK ACROSS THE PFO INTO THE RIGHT ATRIUM. THE PATIENT EXPRESSED DISCOMFORT AND THEN FELT FAINT AND WAS WITHOUT A BLOOD PRESSURE. A TRANS-ESOPHAGEAL ECHOCARDIOGRAM WAS PERFORMED REVEALING A PERICARDIAL EFFUSION. A PERICARDIOCENTESIS WAS PERFORMED WHICH STABILIZED THE PATIENT AND THE PROCEDURE CONTINUED WITH THE PATIENT REMAINING STABLE. THERE WERE NO PERFORMANCE ISSUES NOTED WITH ANY SJM DEVICE USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233323 TACTICATH? QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM CARDIAC ABLATION PERCUTANEOUS CATHETER OAE ST. JUDE MEDICAL PN-004 075 4652623

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention AGILIS NXT INTRODUCER| FASTCATH INTRODUCER| INQUIRY OPTIMA CATHETER