FDA Adverse Event
Malfunction
Summary report: N
RADSUITE
MDR report key: 4668279
·
Received April 2, 2015
Report
- Report Number
- 2183926-2015-00001
- Event Type
- Malfunction
- Date Received
- April 2, 2015
- Date of Event
- February 25, 2015
- Report Date
- March 30, 2015
- Manufacturer
- MERGE HEALTHCARE INCORPORATED
- Product Code
- LLZ
- PMA / PMN Number
- K053281
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
RADSUITE IS A MEDICAL IMAGE VIEWING SOFTWARE APPLICATION THAT ALLOWS IMAGE DATA TO BE VIEWED FROM A FACILITY IMAGE CACHE (FIC) OR FROM A LONG TERM ARCHIVE (LTA). A CUSTOMER NOTICED THAT PATIENT A'S STUDY WAS STORED UNDER THE CORRECT PATIENT IN THE FIC BUT WAS STORED AS A DIFFERENT, PATIENT B, IN THE LTA. WHEN PATIENT A'S CURRENT STUDY WAS OPENED IN THE RADSUITE VIEWER ON THE LTA, PATIENT B'S STUDY WAS OPENED AS A PRIOR INSTEAD OF PATIENT A'S STUDY, AND WAS LABELED WITH PATIENT A'S INFORMATION.