PARKER BATH
Report
- Report Number
- 3007420694-2015-00066
- Event Type
- Injury
- Date Received
- April 2, 2015
- Date of Event
- March 23, 2015
- Report Date
- March 23, 2015
- Manufacturer
- ARJO HOSP EQUIPMENT AB
- Product Code
- ILM
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(6). ADD'L INFO WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.
INITIALLY IT WAS REPORTED BY ARJOHUNTLEIGH'S REP THAT: "CLIENT WAS FOUND UNCONSCIOUS IN THE WATER". CAREGIVER FILLED FOOTWELL OF THE BATH, PUT CLIENT INTO BATH AND CONTINUED TO RUN WATER WHILE TILTING BATH BACK SLIGHTLY, CAREGIVER THEN STARTED TO COOK CLIENT'S BREAKFAST, SECOND CAREGIVER WALKED INTO HOME TO FIND CLIENT FACEDOWN IN BATH, SHE IMMEDIATELY CALLED FOR HELP AND BOTH STAFF MEMBERS LIFTED CLIENT OUT OT THE BATH AND STARTED CPR TO RESUSCITATE THE CLIENT, CLIENT CAME AROUND WHERE THEY CALLED PARAMEDICS AND COMFORTED UNTIL THEY ATTENDED. THE PARAMEDICS WERE CALLED AND CONFIRMED SHE WAS BETTER TO STAY IN HER OWN ENVIRONMENT AFTER THEY PASSED HER HEALTH. DEVICE EXAMINATION PERFORMED WITH THE CUSTOMER SHOWED THAT THE INVOLVED BATH IS IN "EXCELLENT CONDITION". BATH OPERATED THROUGH ALL ITS FUNCTIONS PERFECTLY BUT WHEN FILLING AUTOSHOP DEVICE DOES NOT ACTIVATE SO WATER DOES NOT STOP UNTIL MANUALLY OPERATED. THE RESIDENT WAS ASSESSED IN RELATION TO MOBILITY GALLERY AS "B"- PARTLY CAPABLE OF PERFORMING DAILY INDEPENDENTLY AND THE ASSISTANCE SHE REQUIRES IS NOT GENERALLY PHYSICALLY DEMANDING FOR THE CAREGIVER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 216476 | PARKER BATH | ILM | ARJO HOSP EQUIPMENT AB | AL25110-GB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |