FDA Adverse Event Malfunction Summary report: N

TRACER HYBRID WIRE GUIDE

MDR report key: 4651059 · Received March 27, 2015

Report

Report Number
1037905-2015-00123
Event Type
Malfunction
Date Received
March 27, 2015
Date of Event
February 24, 2015
Report Date
February 24, 2015
Manufacturer
WILSON-COOK MEDICAL INC.
Product Code
KOG
PMA / PMN Number
K910497
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: OUR EVALUATION OF THE RETURNED DEVICE CONFIRMED THE REPORT. THE WIRE GUIDE WAS RETURNED STILL INSIDE THE FS-QEB-A BALLOON. THE WIRE GUIDE WENT IN THE IDE PORT AND OUT THE DISTAL END. APPROXIMATELY 9 CM FROM THE DISTAL END IS A 7 CM SECTION OF EXPOSED CORE WIRE. DUE TO THE CONDITION OF THE COATING IT CANNOT BE DETERMINED IF A SECTION IS MISSING. THE LENGTH OF THE RETURNED WIRE GUIDE IS 186.5 CM. THE STANDARD LENGTH OF THE WIRE GUIDE IS 480 CM. APPROXIMATELY 293.5 CM OF THE WIRE GUIDE WAS NOT WITH THE RETURN AND APPEARS TO HAVE BEEN CUT OFF. THE IDE PORT OF THE FS-QEB-A BALLOON WAS MEASURED AND THE SMALLEST MEASUREMENT IS WITHIN THE SPECIFICATION. THE BALLOON WAS ADVANCED OVER A NEW WIRE GUIDE WITH THE WIRE GUIDE EXITING THE IDE PORT. THE WIRE GUIDE HAD BEEN ADVANCED INTO AN OLYMPUS JF 130 3.2 SIDE VIEWING ENDOSCOPE IN A SIMULATED UPPER GL POSITION WITH THE TIP IN A RETROFLEXED POSITION TO SIMULATE A WORST CASE SITUATION. WHEN THE BALLOON WAS RETRACTED FROM THE ENDOSCOPE OVER THE NEW WIRE GUIDE NO DAMAGE WAS OBSERVED TO THE WIRE GUIDE. A PRODUCT-SPECIFIC DISCREPANCY THAT COULD HAVE CAUSED OR CONTRIBUTED TO THIS OBSERVATION WAS NOT OBSERVED DURING OUR LABORATORY ANALYSIS. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL USE CONDITIONS COULD NOT BE DUPLICATED IN THE LABORATORY SETTING. DUE TO A VARIETY OF CLINICAL CONDITIONS SUCH AS PATIENT ANATOMY, ENDOSCOPE POSITION OR PROGRESSION OF DISEASE STATE, WE COULD NOT REPRODUCE THE ACTUAL CONDITIONS OF PRODUCT USAGE DURING OUR LABORATORY ANALYSIS. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT LINE INSTRUCT THE USER TO FLUSH THE WIRE GUIDE PRIOR TO REMOVAL FROM THE WIRE GUIDE HOLDER. THIS ACTIVITY WILL AID IN OPTIMAL PERFORMANCE OF THE WIRE GUIDE. FAILURE TO FLUSH THE WIRE GUIDE CAN RESULT IN DAMAGE TO THE WIRE GUIDE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT LINE INSTRUCT THE USER TO FLUSH ENDOSCOPE ACCESSORY CHANNEL AND/OR LUMEN OF DEVICE WITH STERILE WATER AND FOR BEST RESULTS, WIRE GUIDE SHOULD BE KEPT WET. THIS ACTIVITY WILL AID IN OPTIMAL PERFORMANCE OF THE WIRE GUIDE. FAILURE TO FLUSH THE ENDOSCOPE CHANNEL CAN RESULT IN DAMAGE TO THE WIRE GUIDE. THE INSTRUCTIONS FOR USE FOR THIS PRODUCT CAUTIONS THE USER THAT THIS PRODUCT IS COMPATIBLE WITH NON-METAL TIP DEVICES. USE OF THE WIRE GUIDE WITH METAL TIP DEVICES MAY COMPROMISE THE INTEGRITY OF THE EXTERNAL COATING ON THE WIRE GUIDE. THE REPORTED OBSERVATION CAN OCCUR IF THE WIRE GUIDE WAS USED WITH AN INCOMPATIBLE ACCESSORY DEVICE. IF ADDITIONAL PRESSURE IS APPLIED TO THE WIRE GUIDE AND/OR ACCESSORY DEVICE(S) WHILE MOVING THE WIRE GUIDE INSIDE THE ACCESSORY DEVICE(S), THIS COULD CONTRIBUTE TO WIRE GUIDE COATING DAMAGE. PRIOR TO DISTRIBUTION, ALL TRACER HYBRID WIRE GUIDES ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MANUFACTURING REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP), THE PHYSICIAN USED A COOK TRACER HYBRID WIRE GUIDE. THE BALLOON CATHETER WAS ADVANCED OVER THE PREPOSITIONED WIRE GUIDE WHICH WAS IN THE BILE DUCT AS THE BALLOON WAS BEING ADVANCED OVER THE WIRE GUIDE UP INTO THE BILE DUCT, IT STRIPPED OFF THE FLOPPY COATING ON THE DISTAL END OF THE WIRE GUIDE. NO SECTION OF THE DEVICE DETACHED AND BECAME FREE INSIDE THE ENDOSCOPE OR PATIENT A NEW BALLOON AND WIRE GUIDE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PATIENT'S BODY. THE PATIENT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
205169 TRACER HYBRID WIRE GUIDE KOG, ENDOSCOPE AND/OR ACCESSORIES KOG WILSON-COOK MEDICAL INC. W3529901

Patients

Seq Age Sex Outcome Treatment
1 COOK FUSION MONI-TOME SPHINTEROTOME (FS-OMNI)| COOK FUSION QUATTRO EXTRACTION BALLOON (FS-QEB-A)| OLYMPUS TJF-Q180V ENDOSCOPE