FDA Adverse Event Malfunction Summary report: N

ELITE CARE CLINER W/SWING AWAY ARMS

MDR report key: 4647435 · Received March 30, 2015

Report

Report Number
1027229-2015-00001
Event Type
Malfunction
Date Received
March 30, 2015
Date of Event
December 31, 2014
Report Date
February 27, 2015
Manufacturer
WINCO MFG LLC
Product Code
FRJ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FACILITY CONTACTED WINCO ABOUT THE INCIDENT, BY PHONE, ON 12/31/2014 BUT REFUSED TO RETURN THE CHAIR TO WINCO FOR EVAL UNTIL 01/29/2015. THIS IS THE "BECOMES AWARE" DATE THAT WE USED. ON 02/27/2015, ADVISED CLINIC TO INSPECT THEIR REMAINING RECLINERS FOR WORN OR INCORRECTLY INSTALLED WIRING HARNESS.

Description of Event or Problem · 1

ON (B)(6) 2014 WINCO CUSTOMER SVC REC'D A CALL FROM (B)(6), AREA TECH OPERATIONS MGR FOR FRESENIUS MEDICAL CARE. HE INFORMED US THAT A 6940 ELITE CARE CLINER HAD OVERHEATED DURING A PT TREATMENT AND LEFT A BURN PATTERN ON THE CHAIR BACK CUSHION AND THE PTS PILLOW. THE PT INITIALLY HAD RED MARKS ON HIS BACK SIMILAR TO A SUNBURN. THE PT REFUSED TO GO TO THE HOSP. UPON COMPLETION OF DIALYSIS, THE PT HAD ONLY A SMALL RED MARK ON HIS BACK. PT REC'D NO MEDICAL CARE FOR THE RED MARKS (B)(4).THE FACILITY REFUSED REPEATED REQUESTS FROM WINCO TO RETURN THE RECLINER FOR EVAL FOR APPROXIMATELY ONE MONTH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
208319 ELITE CARE CLINER W/SWING AWAY ARMS CLINICAL RECLINER, 6940-19-TB-HM FRJ WINCO MFG LLC 6940 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK