FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4646854 · Received March 31, 2015

Report

Report Number
2531779-2015-10632
Event Type
Malfunction
Date Received
March 31, 2015
Report Date
March 17, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 05/13/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED A MANUAL TIME/DATE CHANGE FROM (B)(4) 2015 21:39 TO (B)(4) 2015 21:39. SEVERAL MANUAL TIME/DATE CHANGES WERE OBSERVED IN THE BLACK BOX BETWEEN (B)(4) 2015 AND (B)(4) 2015. NO ALARMS OCCURRED DURING THE INVESTIGATION AND THE PUMP PASSED A TIME TEST. THE PUMP WAS ABLE TO MAINTAIN THE TIME/DATE SETTINGS WITH ACCURACY.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE TIME AND DATE OF THE PUMP WAS SWITCHING BACK AND FORTH BETWEEN TWO DAYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
212015 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR