FDA Adverse Event Injury Summary report: N

POLYAXIAL DRIVER, REFORM

MDR report key: 4634913 · Received March 24, 2015

Report

Report Number
3005739886-2015-00034
Event Type
Injury
Date Received
March 24, 2015
Date of Event
December 18, 2014
Report Date
March 2, 2015
Manufacturer
PRECISION SPINE, INC.
Product Code
HXX
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ENGINEERING EVALUATION OF THE RETURNED DRIVER CONCLUDED THAT THE ROOT CAUSE OF THE FRACTURED TIP IS ATTRIBUTED TO THE DRIVER NOT BEING FULLY SEATED IN THE POLYAXIAL SCREW. THE PLASTIC OUTER SLEEVE OF THE ASSEMBLY WAS PREVIOUSLY MODIFIED TO MITIGATE THE OCCURRENCE OF THE USER UNTHREADING THE OUTER SECURE SHAFT FROM THE TULIP. THIS CHANGE WAS IMPLEMENTED IN MARCH OF 2014. ADDITIONALLY, A DESIGN CHANGE ON THE THREAD-FORM OF THE POLYAXIAL SCREW, HAS BEEN IMPLEMENTED TO REDUCE THE AMOUNT OF THREAD-WASHOUT, THUS REDUCING THE INSERTION TORQUE, TO MITIGATE FAILURES DUE TO THE AFOREMENTIONED CONDITION. REVIEW OF MANUFACTURING HISTORY FOUND A TOTAL OF (B)(4) PIECES OF THIS LOT WERE RELEASED FOR DISTRIBUTION ON NOVEMBER 8, 2014 WITH NO DEVIATION OR ANOMALIES. THIS LOT WAS MANUFACTURED TO A PREVIOUS REVISION OF THE PLASTIC OUTER SLEEVE. DESIGN IMPROVEMENTS HAVE BEEN INITIATED FOR THE SHAFT SLEEVE COMPONENT SUBSEQUENT TO THE MANUFACTURE OF THIS LOT. THIS REPORT IS NUMBER 1 OF 3 MDR'S FILED FOR THE SAME EVENT (REFERENCE 30057398896-2015-00034/36).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A L3-S1 PROCEDURE PERFORMED ON (B)(6) 2014, THE TIP OF TWO REFORM POLYAXIAL DRIVERS (39-SP-0700) BROKE WHILE IMPLANTING THE 7.5 X 45 MM REFORM POLYAXIAL SCREWS. THE SCREWS WERE REMOVED AND REPLACED. NO DELAY TO THE PROCEDURE RESULTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
198071 POLYAXIAL DRIVER, REFORM SCREWDRIVER HXX PRECISION SPINE, INC. 3144MMR

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention