FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4633695 · Received March 25, 2015

Report

Report Number
2531779-2015-09976
Event Type
Malfunction
Date Received
March 25, 2015
Report Date
March 13, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 05/12/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: THE BLACK BOX SHOWED A MANUAL DATE CHANGE WAS MADE FROM (B)(4) 2015 18:18 TO (B)(4) 2015 18:18. AND A MANUAL TIME WAS MADE FROM (B)(4) 2015 23:55 TO (B)(4) 2015 00:55. NO ALARMS OCCURRED DURING THE INVESTIGATION AND THE PUMP PASSED A TIME TEST. THE PUMP IS ABLE TO MAINTAIN THE TIME/DATE SETTINGS WITH ACCURACY. UNRELATED TO THE INITIAL ALLEGATION, THE DISPLAY SCREEN WAS DIM AND DISCOLORED.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE TIME AND DATE ON THE PUMP HAD BECOME INCORRECT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201199 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR