FDA Adverse Event Injury Summary report: N

TRIDENT ALUMINA INSERT

MDR report key: 4632703 · Received March 25, 2015

Report

Report Number
0002249697-2015-00898
Event Type
Injury
Date Received
March 25, 2015
Date of Event
January 1, 2013
Report Date
March 2, 2015
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MRA
PMA / PMN Number
P000013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING MALPOSITION INVOLVING BOTH AN UNKNOWN OMNIFIT STEM AND UNKNOWN TRIDENT SHELL WAS REPORTED. THE MATERIAL ANALYSIS CONCLUDED: "BOTH THE CERAMIC HEAD AND INSERT/SLEEVE INDICATED EVIDENCE OF WEAR SCARRING AND EXPLANTATION DAMAGE. MATERIAL DAMAGE LIKELY DUE TO IMPINGEMENT WAS OBSERVED ON THE DISTAL SLEEVE RIM. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON ANY OF THE COMPONENTS EXAMINED." A REVIEW OF THE DEVICE HISTORY RECORDS INDICATES THAT THE REPORTED DEVICES WERE MANUFACTURED AND ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE COMPLAINT HISTORY REVIEW INDICATED THAT THERE WERE NO SIMILAR EVENTS FOR THE REPORTED LOT. CONCLUSION: BASED ON THE PROVIDED INFORMATION AND THE MEDICAL PERFORMED THE PRODUCT REPORTED IN THIS INVESTIGATION DID NOT CONTRIBUTE TO THE EVENT. ROOT CAUSE OF FAILURE IN THIS CASE IS RELATED TO COMPONENT MALPOSITION OF BOTH THE STEM AND THE SHELL. NO FURTHER INVESTIGATION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE TO INDICATE FURTHER EVALUATION IS WARRANTED, THIS RECORD WILL BE REOPENED.

Description of Event or Problem · 1

THE PATIENT UNDERWENT THA DUE TO OA WITH CERAMIC ON CERAMIC FOR RIGHT HIP ON (B)(6) 2008. OSTEOLYSIS WAS NOTICED ON 2013 AT FEMORAL SIDE AND ACETABULAR SIDE. (B)(6) 2015, ALUMINA CERAMIC CUP AND ALUMINA CERAMIC LINER WERE REPLACED TO X3 POLY LINER AND DELTA CERAMIC.

Description of Event or Problem · 1

THE PATIENT UNDERWENT THA DUE TO OA WITH CERAMIC ON CERAMIC FOR RIGHT HIP ON (B)(6)2008. OSTEOLYSIS WAS NOTICED ON 2013 AT FEMORAL SIDE AND ACETABULAR SIDE. (B)(6) 2015, ALUMINA CERAMIC CUP AND ALUMINA CERAMIC LINER WERE REPLACED TO X3 POLY LINER AND DELTA CERAMIC.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
201091 TRIDENT ALUMINA INSERT IMPLANT MRA STRYKER ORTHOPAEDICS-MAHWAH 24138601

Patients

Seq Age Sex Outcome Treatment
1 74 YR Hospitalization| R