FDA Adverse Event
Malfunction
Summary report: N
IMC PISTON SYRINGE
MDR report key: 4631286
·
Received March 20, 2015
Report
- Report Number
- 1423625-2015-00001
- Event Type
- Malfunction
- Date Received
- March 20, 2015
- Date of Event
- February 1, 2015
- Report Date
- February 19, 2015
- Manufacturer
- INTL. MEDSURG CONNECTION, INC.
- Product Code
- FMF
- PMA / PMN Number
- K102969
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
A 60CC SYRINGE SPLIT WHILE DOING A BEDSIDE PERICENTESIS. THE RESIDENT WAS SPLASHED IN THE FACE WITH THE FLUID. NO INJURY TO THE RESIDENT WAS REPORTED. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO PATIENT IMPACT. EVALUATION OF THE RETURNED SAMPLE CONFIRMED A CRACK IN THE BARREL OF THE SYRINGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 191748 | IMC PISTON SYRINGE | SYRINGE | FMF | INTL. MEDSURG CONNECTION, INC. | 300-60LL | 13-7818-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |