FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4629154 · Received March 24, 2015

Report

Report Number
2531779-2015-09816
Event Type
Malfunction
Date Received
March 24, 2015
Report Date
March 9, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/09/2015 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS PEELING FROM THE BASE. ALL OF THE KEYPAD BUTTONS WERE RESPONSIVE DURING THE INVESTIGATION. CONTAMINATION WAS FOUND UNDERNEATH THE KEYPAD CONTACTS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE KEYPAD BUTTON CONTACTS HAD CONTAMINATION. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 03/09/2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
196073 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 18 YR