FDA Adverse Event Injury Summary report: N

MELODY TRANSCATHER PULMONARY

MDR report key: 4624842 · Received March 23, 2015

Report

Report Number
2025587-2015-00352
Event Type
Injury
Date Received
March 23, 2015
Date of Event
December 3, 2014
Report Date
April 22, 2015
Manufacturer
HEART VALVES SANTA ANA
Product Code
NPV
PMA / PMN Number
P140017
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCTS WERE NOT RETURNED TO MEDTRONIC. (B)(4). TITLE: PERCUTANEOUS PULMONARY VALVE IMPLANTATION: 5 YEARS OF FOLLOW-UP. DOES AGE INFLUENCE OUTCOMES? AUTHORS: SHARON BORIK, MD; ANDREW CREAN, MD, FRCPC; ERIC HORLICK, MDCM, FRCPC; MARK OSTEN, MD, FRCPC; KYONG-JIN LEE, MD, FRCPC; RAJIV CHATURVEDI, MD, PHD, MRCP(UK); MARK K. FRIEDBERG, MD; BRIAN W. MCCRINDLE, MD, FRCPC; CEDRIC MANLHIOT, BSC; LEE BENSON, MD, FRCPC JOURNAL CITATION: CIRCULATION CARDIOVASCULAR INTERVENTIONS; 2015;8:E001745 DOI: 10.1161/CIRCINTERVENTIONS.114.00174.

Additional Manufacturer Narrative · 1

SUPPLEMENTAL SUBMITTED TO ATTACH LITERATURE; NO OTHER ADDITIONS OR CHANGES.

Additional Manufacturer Narrative · 1

CONCLUSION: THE STERILIZATION PROCESS USED BY MEDTRONIC HAS AN ANTI-MICROBIAL KILL ON THE MICROORGANISMS SUCH AS STAPHYLOCOCCUS AND STREPTOCOCCUS SPECIES. THE REPORTED ORGANISM CAN BE RENDERED NON-VIABLE TO THE STERILIZATION PROCESS DURING MANUFACTURING. THIS PROCESS IS VALIDATED USING THE WORST CASE BACILLUS ATROPHAUS (GRAM POSITIVE SPORE FORMING RODS), WHICH IS KNOWN TO BE REPRESENTATIVE OF THE CLEANROOM BIOBURDEN AND OUR PROCESS IS VALIDATED IS EFFECTIVE. ENDOCARDITIS THAT OCCURS MORE THAN 12 MONTH AFTER THE PROCEDURE ARE CALLED LATE PROSTHETIC-VALVE ENDOCARDITIS AND ARE LARGELY COMMUNITY-ACQUIRED VERSUS A RESULT OF THE MANUFACTURING PROCESS OF THE VALVE. BASED ON CLINICAL DATA AND LITERATURES, MELODY STENT FRACTURES ARE KNOWN PHENOMENON. PROMINENT MECHANICAL STRESSES ON THE OUTFLOW TRACT STENT, SUCH AS COMPRESSION BETWEEN THE ANTERIOR CHEST WALL AND HEART, APPEAR TO BE ASSOCIATED WITH AN INCREASED RISK OF STENT FRACTURE. HOWEVER, A TRUE ROOT CAUSE TO THE STENT FRACTURES WAS NOT DETERMINED. OVERALL, A TRUE ROOT CAUSE TO THE REPORTED CLINICAL OBSERVATIONS WAS NOT DETERMINED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT A STUDY WAS PERFORMED TO ASSESS THE PERSISTENCE OF REVERSE RIGHT VENTRICLE (RV) REMODELING FOLLOWING PERCUTANEOUS PULMONARY VALVE IMPLANTATION (PPVI) AS WELL AS THE TIMING PPVI INFLUENCED OUTCOMES. THE STUDY INCLUDED 51 PATIENTS (INCLUDING 23 PATIENTS <(><<)>16 YEARS OLD) WHO UNDERWENT IMPLANTATION OF A MEDTRONIC BIOPROSTHETIC PULMONARY VALVE AND WERE FOLLOWED FOR A MEAN OF 4.5 YEARS POST-IMPLANT BETWEEN OCTOBER 2005 AND OCTOBER 2011. THE PATIENT POPULATION WAS PREDOMINANTLY MALE WITH A MEAN AGE OF 20.2 (±10.8) YEARS OLD, AND A MEAN WEIGHT OF 58.3 (±16.0) KILOGRAMS. FREEDOM FROM ANY RE-INTERVENTION WAS 87% AND 68% AT 3 AND 5 YEARS RESPECTIVELY, AND FREEDOM FROM SURGERY WAS 90% AT 5 YEARS. THERE WERE FIVE OPEN CHEST SURGICAL INTERVENTIONS, WHICH INCLUDED RV TO PULMONARY ARTERY CONDUIT REPLACEMENTS (N=2), PULMONARY VALVE REPLACEMENT (N=1), ARTERIAL SWITCH OPERATION (N=1), AND AN AORTIC VALVE AND ROOT REPLACEMENT WITH PULMONARY HOMOGRAFT IMPLANTATION FOR INFECTIOUS ENDOCARDITIS (N=1). THE PATIENT WITH ENDOCARDITIS PRESENTED FIVE YEARS POST-PPVI WITH GRAM-POSITIVE COCCI ON THE IMPLANTED VALVE. THERE WERE SEVEN PERCUTANEOUS SURGICAL INTERVENTIONS, WHICH INCLUDED RE-VALVING (NEW BIOPROSTHETIC VALVE IMPLANTED WITHIN ORIGINAL BIOPROSTHETIC VALVE; N=5), AND BALLOON VALVULOPLASTY DILATIONS (N=2) TO ADDRESS STENOSIS OR RV HYPERTENSION. ALSO, 18% OF PATIENTS (N=9) HAD EVIDENCE OF MINOR TYPE 1 STENT FRACTURES VIA DIGITAL CHEST RADIOGRAPHY, AND 3% OF PATIENTS (N=1) EXHIBITED STENT FRACTURE WITH LOSS OF STENT INTEGRITY WHICH REQUIRED RE-VALVING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
192810 MELODY TRANSCATHER PULMONARY PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED NPV HEART VALVES SANTA ANA PB1018

Patients

Seq Age Sex Outcome Treatment
1 00020 YR Life Threatening| R