LEICA M320 W12
Report
- Report Number
- 3003974370-2015-00001
- Event Type
- Malfunction
- Date Received
- March 18, 2015
- Date of Event
- February 6, 2014
- Report Date
- February 6, 2014
- Manufacturer
- LEICA MICROSYSTEMS (SCHWEIZ) AG
- Product Code
- EPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE MFR REPORT NUMBER OF THE INITIAL REPORT SUBMITTED ON DATE 02/20/2015 WAS ERRONEOUSLY SHOWN AS 3003974370-2015-00016. THE MFR REPORT NUMBER FOR THE EVENT AT DR. (B)(6) FACILITY IS IDENTIFIED AS 3003974370-2015-00001 AND HAD BEEN CORRECTED IN SECTION "MFT REPORT#". THIS IS A FINAL REPORT. AN INVESTIGATION ON THE AFFECTED PART OF THE SURGICAL MICROSCOPE (PARALLELOGRAM) WAS CONDUCTED BY THE SPECIFICATION DEVELOPER. VISUAL INSPECTION AND FUNCTIONAL EVAL OF THE AFFECTED COMPONENT WERE PERFORMED. THE MALFUNCTION WAS NOT REPRODUCIBLE TO RESULT IN THE SAME DAMAGES OF SCREWS AS REPORTED. IT WAS FOUND THAT THE ROOT CAUSE IS THAT THE SCREWS OF A COMPONENT WERE NOT PROPERLY TIGHTENED DUE TO HUMAN ERROR AT THE SUPPLIER'S MFG SITE BASED ON INVESTIGATION AT THIS SUPPLIER. THERE ARE NO INDICATIONS THAT THE PROBLEM IS SYSTEMATIC. BASED ON THE RESULTS, THE PROBABILITY OF OCCURRENCE IS REMOTE. THEREFORE THE INCIDENT CAN BE CONSIDERED AS AN ISOLATED EVENT.
(B)(4). ON 02/06/2015, LEICA MICROSYSTEMS RECEIVED A COMPLAINT STATING THAT THE SWING ARM OF THE LEICA M320 W12 SURGICAL MICROSCOPE FELL TO THE LOWEST ARM POSITION BESIDE THE PT. THE SURGEON WAS ABLE TO HOLD ON TO ONE OF THE EYE PIECES TO STOP THE FALL. IT WAS REPORTED THAT THERE WAS NO INJURY TO PT OR USER. THE SURGICAL MICROSCOPE IS OUT OF SERVICE AND THE AFFECTED PART IS BEING SHIPPED TO MANUFACTURER FOR INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 183669 | LEICA M320 W12 | MICROSCOPE, SURGICAL | EPT | LEICA MICROSYSTEMS (SCHWEIZ) AG | M320 W12 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |