FDA Adverse Event Malfunction Summary report: N

HEAD RING POST

MDR report key: 4607903 · Received March 11, 2015

Report

Report Number
1222895-2015-00014
Event Type
Malfunction
Date Received
March 11, 2015
Report Date
January 27, 2015
Manufacturer
INTEGRA BURLINGTON, MA, INC.
Product Code
HAW
PMA / PMN Number
K944463
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVAL. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFO.

Additional Manufacturer Narrative · 1

INTEGRA COMPLETED ITS INTERNAL INVESTIGATION ON 06/08/2015. RESULTS: DEVICE HISTORY RECORD REVIEWED FOR LOT #0308297 MANUFACTURED ON MAY 6, 2014 SHOW NO ANOMALIES RELATED TO THE REPORTED FAILURE. THIS DEVICE PASSED ALL REQUIRED INSPECTION POINTS WITH NO ASSOCIATED MRR'S, VARIANCES OR REWORK. TWO YEAR LOOKBACK IN COMPLAINT MANAGEMENT SYSTEM FOR REPORTED FAILURE AND OR RELATED TO "EDGING WAS SHARP AND NOTCHES IN SOME" FOR THIS PRODUCT ID SHOWS THAT 3 COMPLAINTS WERE RECEIVED INCLUDING THIS CASE. ALL THREE COMPLAINTS WERE RECEIVED ON THE SAME DATE BY THE SAME CUSTOMER. NO NEW DESIGN OR MFG TRENDS HAVE BEEN IDENTIFIED. COMPLAINT NOT CONFIRMED, THERE WERE NO RAISED EDGES OR SHARP EDGES ON RETURNED PRODUCT. CONCLUSION: NO ROOT CAUSE ANALYSIS COULD BE PERFORMED AS THE COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

THIS IS THE FIRST REPORT OF THREE INVOLVING THE SAME PRODUCT. CROSS REFERENCE WITH MFR. REPORT NUMBER 1222895-2015-00012 (PR 124673) AND MFR. REPORT NUMBER 1222895-2015-00013 (B)(4). IT WAS REPORTED THAT A HRP- HEAD RING POST; HAS EDGING THAT WAS SHARP WITH NOTCHES. THERE WAS NO PT INVOLVEMENT WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
167363 HEAD RING POST NA HAW INTEGRA BURLINGTON, MA, INC. 0308297

Patients

Seq Age Sex Outcome Treatment
1