FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT

MDR report key: 4602217 · Received March 15, 2015

Report

Report Number
9611451-2015-00132
Event Type
Malfunction
Date Received
March 15, 2015
Report Date
February 11, 2015
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K122432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT380 EVAQUA2 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND WAS VISUALLY INSPECTED. RESULT: THE VISUAL INSPECTION FOUND THAT THERE WAS NO DAMAGE TO THE MDI PORT OR THE PORT CAP. THE MDI PORT AND PORT CAP WERE MEASURED AND WERE FOUND TO BE OUT OF SPECIFICATION. A LOT CHECK WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND FOR LOT 140909. CONCLUSION: WE WERE UNABLE TO DETERMINE THE CAUSE OF THE REPORTED EVENT. ALL RT380 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE PRODUCT WAS RELEASED FOR DISTRIBUTION, DURING TRANSPORTATION OR STORAGE. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT380 BREATHING CIRCUIT STATE THE FOLLOWING: -"CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." -"SET APPROPRIATE VENTILATOR ALARMS." -"PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN THE (B)(6) REPORTED THAT THE CAP OF THE MDI PORT OF AN RT380 BREATHING CIRCUIT IS DISCONNECTING EASILY.NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
176659 ADULT DUAL HEATED EVAQUA2 BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT380 140909

Patients

Seq Age Sex Outcome Treatment
1