ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2015-08270
- Event Type
- Malfunction
- Date Received
- March 12, 2015
- Report Date
- February 23, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
DATE OF SUBMISSION 03/12/2015-DEVICE EVALUATION: THE RETAINED CARTRIDGE HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON 03/12/2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.
FOLLOW-UP #1 DATE OF SUBMISSION 03/26/2015-DEVICE EVALUATION: THE RETAINED CARTRIDGE HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON 03/12/2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.
FOLLOW-UP #1 DATE OF SUBMISSION 07/16/2015-DEVICE EVALUATION: THE CARTRIDGE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.
THE CORRECT LOT NUMBER FOR THE CARTRIDGE IS C200011.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 172537 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | C200008 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |