FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 4596333 · Received March 12, 2015

Report

Report Number
2531779-2015-08270
Event Type
Malfunction
Date Received
March 12, 2015
Report Date
February 23, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DATE OF SUBMISSION 03/12/2015-DEVICE EVALUATION: THE RETAINED CARTRIDGE HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON 03/12/2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 03/26/2015-DEVICE EVALUATION: THE RETAINED CARTRIDGE HAS BEEN EVALUATED BY PRODUCT ANALYSIS ON 03/12/2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE RETAINED CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 07/16/2015-DEVICE EVALUATION: THE CARTRIDGE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE CARTRIDGE PASSED VISUAL INSPECTION WITH NO DAMAGE OR DEFECTS NOTED. A FILL TEST WAS COMPLETED WITH NO AIR BUBBLES BEING FORMED INSIDE THE CARTRIDGE. A LEAK TEST WAS PERFORMED WITH NO FAILURES BEING OBSERVED; NO LEAKS WERE OBSERVED FROM THE LUER CONNECTION, O-RINGS, OR ANYWHERE ELSE IN THE CARTRIDGE. THERE WAS NO DEFECT FOUND.

Additional Manufacturer Narrative · 1

THE CORRECT LOT NUMBER FOR THE CARTRIDGE IS C200011.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172537 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION C200008

Patients

Seq Age Sex Outcome Treatment
1 28 YR