FDA Adverse Event Malfunction Summary report: N

PLM A+PMP W/BSE MOD

MDR report key: 4595097 · Received March 10, 2015

Report

Report Number
9615050-2015-00596
Event Type
Malfunction
Date Received
March 10, 2015
Date of Event
January 1, 2014
Report Date
December 30, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K011442
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, IT WAS NOTED THAT THE DEVICE DID NOT PASS THE UNRESTRICTED FLOW TEST AND IT WAS FOUND THAT WAS FOUND THAT THE REGULATOR CLOSER WAS NOT SEATED PROPERLY. WHEN THE DEVICE DOOR IS CLOSED, THE REGULATOR OPENER OPENS THE CASSETTE FLOW REGULATOR CLOSER CLOSES THE CASSETTE FLOW REGULATOR. THE PROBABLE CAUSE FOR THE REGULATOR CLOSER TO BE NOT SEATED PROPERLY IS THAT DURING THE INSTALLATION OF FLUID SHIELD, IF THE FLUID SHIELD TAB IS NOT ALIGNED PROPERLY, IT WILL PUSH THE REGULATOR CLOSER AND CAUSE THE REGULATOR CLOSER TO DISENGAGE FROM THE CHASSIS POST. UPON CLOSING THE DOOR LEVER, THE REGULATOR CLOSER WILL DISLODGE AND MOVE TO THE FRONT OF THE CHASSIS POST. AS THE REGULATOR CLOSER IS NOT SNAPPED INTO POSITION ON ITS MATING POST ON THE MECHANISM CHASSIS, THE CASSETTE FLOW REGULATOR WILL NOT BE CLOSED RESULTING IN FREE FLOW WHEN THE DEVICE DOOR IS OPENED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED TO THE SERVICE CENTER FOR A REPORT FROM THE CUSTOMER CONTACT THAT THE DISTAL PRESSURE WAS OUT OF TOLERANCE AND THE DEVICE MECHANISM COULD NOT BE CALIBRATED. THIS IS NOT A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE DEVICE DID NOT PASS THE UNRESTRICTED FLOW TEST AND IT WAS FOUND THAT THE REGULATOR CLOSER WAS NOT SEATED PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166056 PLM A+PMP W/BSE MOD 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA