FDA Adverse Event Malfunction Summary report: N

ISCREEN OFD SALIVA TEST

MDR report key: 4583938 · Received March 9, 2015

Report

Report Number
2027969-2015-00167
Event Type
Malfunction
Date Received
March 9, 2015
Date of Event
February 13, 2015
Report Date
February 13, 2015
Manufacturer
ABON BIOPHARM (HANGZHOU) CO., LTD
Product Code
MVO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NOTE: THIS PRODUCT IS FOR FORENSIC USE ONLY AND THEREFORE IS 510(K) EXEMPT. INVESTIGATION: CUSTOMER'S OBSERVATION COULD NOT BE REPRODUCED WITH IN-HOUSE RETAIN PRODUCT. IN-HOUSE RETAIN PRODUCT WERE TESTED WITH DRUG FREE DONORS, THE TEST RESULTS DID NOT OBSERVED ANY ABNORMAL, INCLUDING SPONGE COLOR, SMELL, DEVICE APPEARANCE ETC. PER ISSUE, "DONOR IS ALLERGIC TO CORN SYRUP", DUE TO SPONGE'S TREATMENT PROCEDURES, IT DID NOT HAVE CORN SYRUP COMPONENT IN SPONGE. SPONGE WAS TREATED WITH SUCROSE WHICH SHARES A COMMON COMPONENT WITH CORN SYRUP. BOTH CORN SYRUP AND SUCROSE CONTAIN GLUCOSE. IT CANNOT BE RULED OUT THAT THIS INGREDIENT MAY HAVE CAUSED THE PATIENT'S ALLERGIC REACTION. THIS ISSUE WILL BE TRACKED AND TRENDED. CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER ALLEGED AN ALLERGIC REACTION FROM USING THE ISCREEN OFD SALIVA TEST. RESULTS AS FOLLOWS: THE DISTRIBUTOR REPORTED THAT THE END-USER HAD A DONOR THAT HAD AN ALLERGIC REACTION AFTER USING THE ISCREEN OFD SALIVA TEST. THE DONOR REPORTS THAT HE IS ALLERGIC TO CORN SYRUP. THE DONOR HAD GONE HOME AFTER USING THE ORAL-FLUID TEST AND CALLED THE TEST PROVIDER BACK AFTER ABOUT 30 MINUTES BECAUSE HIS FACE STARTED TO SWELL. THE DONOR ASKED IF THE TEST COLLECTORS HAD CORN SYRUP IN THEM. THE DONOR WAS REPORTED TO HAVE TAKEN BENADRYL AND WAS DOING OKAY. KIT WAS PURCHASED IN (B)(6) 2014. NO OTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163031 ISCREEN OFD SALIVA TEST DRUGS OF ABUSE TEST MVO ABON BIOPHARM (HANGZHOU) CO., LTD DSB-765-011-ONW510 DOA3090278

Patients

Seq Age Sex Outcome Treatment
1