FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4583368 · Received March 9, 2015

Report

Report Number
2032227-2015-09204
Event Type
Malfunction
Date Received
March 9, 2015
Date of Event
January 4, 2015
Report Date
July 31, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN PROVIDED BY FAILURE ANALYSIS THAT WAS NOT INCLUDED ON THE INITIAL DEVICE EVALUATION: THE MOTOR PASSED THE MOTOR TEST. DJC - SUPPLEMENTAL REPORT WAS CREATED FOR DVA RESULTS.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. HOWEVER, THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT A DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. CUSTOMER STATES THAT THE BLOOD GLUCOSE READING IS 45 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163203 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 58 YR