FDA Adverse Event Malfunction Summary report: N

UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM

MDR report key: 4565850 · Received March 3, 2015

Report

Report Number
1061932-2015-00372
Event Type
Malfunction
Date Received
March 3, 2015
Date of Event
February 4, 2015
Report Date
February 4, 2015
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K140911
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT ON (B)(6) 2015. THE FSE FOUND A CLEANER LEAK DURING SHUTDOWN. THE LEAK WAS CAUSED BY A MISALIGNED PROBE WASH COLLAR. THE FSE PERFORMED PROBE WASH COLLAR ALIGNMENT AND THE INSTRUMENT RAN WITHOUT LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED SERVICE PROCEDURES. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CLEANER FLUID LEAK OF APPROXIMATELY 2ML FROM UNDER THE PROBE ON A UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT AND WAS OBSERVED WHILE THE CUSTOMER WAS PERFORMING PREVENTIVE MAINTENANCE. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT AT THE TIME OF THE EVENT AND THERE WAS NO REPORT OF INJURY OR DIRECT EXPOSURE TO THE LEAK. ERRONEOUS PATIENT RESULTS WERE NOT GENERATED AND THERE WAS NO CHANGE OR AFFECT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT.