FDA Adverse Event
Malfunction
Summary report: N
FREE STANDING INFANT WARMER
MDR report key: 45654
·
Received October 23, 1996
Report
- Report Number
- 2510954-1996-00001
- Event Type
- Malfunction
- Date Received
- October 23, 1996
- Report Date
- September 26, 1996
- Manufacturer
- AIR-SHIELDS
- Product Code
- FMT
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
INFANT RADIANT WARMER IN SERVO (SKIN TEMPERATURE) CONTROL CONTINUED TO HEATER A HIGH SET TEMPERATURE ALARM ACTIVATED. A WARMER DISPLAY TEMPERTURE OF 44 DEGREES CENTIGRADE WAS OBSERVED. IT HAS BEEN INDICATED THAT THE INFANT HAS BEEN RELEASED FROM THE HOSP, WITHOUT THE NEED FOR MEDICAL INTERVENTION, OR SUSTAINING A SERIOUS INJURY RELATING TO THIS EVENT REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREE STANDING INFANT WARMER | INFANT RADIANT WARMER | FMT | AIR-SHIELDS | 7830 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Life Threatening |