FDA Adverse Event Other Summary report: N

VITREA 2

MDR report key: 455386 · Received April 18, 2003

Report

Report Number
1933477-2003-00004
Event Type
Other
Date Received
April 18, 2003
Date of Event
March 21, 2003
Report Date
March 21, 2003
Manufacturer
VITAL IMAGES, INC.
Product Code
LNX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A SITE EXPERIENCED, "DOUBLE PATIENTS IN A REPORT." THE REPORT PAGE FROM VITREA 3.1 HAD ONE PATIENT'S IMAGES WITH A SECOND PATIENT'S NAME IN THE HEADER. THEY CONFIRMED THE SITE HAD USED THE SAME PATIENT'S ID (SINUS) FOR BOTH PATIENTS. AFTER ASCERTAINING THE DETAILS, IT WAS DETERMINED THAT THE REPORT FOR VITREA 3.1 AND 3.2 LOOKS AT THE PATIENT ID ONLY, AND THE DATA FROM THE FIRST PATIENT ID IS CARRIED OVER AND PLACED ON THE SCANNED RESULTS OF ALL OF THE PATIENTS SCANNED WITH THE SAME ID. FURTHER TESTING CONFIRMED THAT THE SLIDE TRAY OPERATES AS DESIGNED, BUT THAT THE HEADER DISPLAYS THE INFORMATION FOR THE FIRST LOADED PATIENTS, RATHER THAN THE LAST LOADED PATIENT. THE CURRENT USING VITREA 3.2 MANUAL STATES, "IF YOU INCLUDE IMAGES FOR MORE THAN ONE SERIES FOR THE SAME PATIENT, IN A REPORT, THE REPORT HEADINGS, IF ANY, WILL IDENTIFY THE SERIES YOU LOADED MOST RECENTLY. IT MAY BE IMPORTANT TO IDENTIFY WHICH IMAGES COME FROM THE SERIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITREA 2 3D RADIOLOGY VIEWER SOFTWARE LNX VITAL IMAGES, INC. VPMC-7299K *

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other