FDA Adverse Event Malfunction Summary report: N

HEARTSINE SAMARITAN 300P AND PAD PAK

MDR report key: 4545643 · Received February 20, 2015

Report

Report Number
3004123209-2015-00178
Event Type
Malfunction
Date Received
February 20, 2015
Date of Event
February 9, 2015
Report Date
August 26, 2015
Manufacturer
HEARTSINE TECHNOLOGIES, LTD
Product Code
MKJ
PMA / PMN Number
K041067
Removal / Correction Number
Z-0124-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE RECORDS 53 LOG ENTRIES BETWEEN APRIL 2007 AND THE 6TH FEBRUARY 2015. THE DEVICE RECORDS 53 MANUAL POWER UPS OF UNDER UP TO 6TH FEBRUARY 2015, 18 OF WHICH LAST TEN MINUTES IN DURATION. NO FURTHER LOG ENTRIES WERE RECORDED. INVESTIGATION FOUND THE REPORTED FAULT CAN BE ATTRIBUTED TO MEMBRANE FAILURE. THE TEN MINUTE TIME OUTS WOULD SUGGEST A FAILING MEMBRANE. THE PAD-PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELED FOR SINGLE USE BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.

Description of Event or Problem · 1

THERE WAS NO PATIENT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE DEVICE WAS SWITCHING ON AUTOMATICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124072 HEARTSINE SAMARITAN 300P AND PAD PAK AUTOMATED EXTERNAL DEFIBRILLATOR MKJ HEARTSINE TECHNOLOGIES, LTD PAD

Patients

Seq Age Sex Outcome Treatment
1