REUSABLE ADULT BREATHING CIRCUIT
Report
- Report Number
- 9611451-2015-00103
- Event Type
- Malfunction
- Date Received
- February 25, 2015
- Date of Event
- January 20, 2015
- Report Date
- January 27, 2015
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K131957
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT FIVE BREATHING CIRCUITS (1 X LOT 140404, MANUFACTURED 4 APRIL 2014; 1 X LOT 140708, MANUFACTURED 8 JULY 2014; 1 X LOT 140721, MANUFACTURED 21 JULY 2014; 2 X LOT UNKNOWN) WERE RECEIVED AT FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. A WATER BATH TEST WAS PERFORMED TO DETECT LEAKS IN THE BREATHING CIRCUITS. THE CIRCUITS WERE VISUALLY INSPECTED, MEASURED FOR BEAD WIDTH, BEAD HEIGHT AND FILM THICKNESS. RESULTS: THE WATER BATH TEST REVEALED THAT ONE OF THE FIVE CIRCUITS WAS LEAKING BETWEEN THE CHAMBER CUFF AND BREATHABLE FILM AND HAD NO OBVIOUS PHYSICAL DAMAGE. VISUAL INSPECTION OF THE OTHER FOUR CIRCUITS REVEALED A TORN FILM IN THE HEATED TUBE. THE SMALLEST TEAR WAS FOUND TO BE APPROXIMATELY 13 MM AND THE BIGGEST TEAR WAS APPROXIMATELY 23 MM. THE TEAR IN THE TUBING FILM WAS FOUND FOR THREE CIRCUITS AT THE PATIENT END OVER-MOULDED CUFF AND FOR ONE CIRCUIT AT THE CHAMBER END OVERMOULDED CUFF. THESE FOUR TORN CIRCUITS HAD A MEASURED BEAD WIDTH AND BEAD HEIGHT WITHIN SPECIFICATION AND ALSO THE FILM THICKNESS WAS WITHIN SPECIFICATION. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT HAD CAUSED THE OBSERVED DAMAGE. THE CUSTOMER HAD REPORTED THAT THE CIRCUITS WERE IN USE FOR ABOUT THREE DAYS AND HAD ALL BEEN PASTEURIZED AND REUSED AT LEAST THREE TIMES, WHICH INDICATES THAT THE CIRCUITS BECAME DAMAGED DURING USE. OUR USER INSTRUCTIONS THAT ACCOMPANY THE (B)(4) REUSABLE BREATHING CIRCUIT CONTAIN THE FOLLOWING WARNINGS: CLEAN CIRCUIT PRIOR TO USE AND AFTER EACH PATIENT USE, USING APPROVED DISINFECTION METHODS ONLY. USE OF UNAPPROVED CLEANING METHODS MAY DAMAGE THE CIRCUIT AND REDUCE IT USEABLE LIFE. INSPECT CIRCUIT BEFORE RE-USE, DO NOT USE IF THE CIRCUIT SHOWS SIGN OF DETERIORATION, SUCH AS CRACKS, TEARS OR DAMAGE. DISCONNECT TUBE BY HANDLING END CONNECTORS ONLY, DO NOT PULL OR TWIST TUBING, AS THIS MAY CAUSE DAMAGE.
(B)(4). THE FIVE COMPLAINT BREATHING CIRCUITS ARE CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT FIVE 900MR810 REUSABLE ADULT BREATHING CIRCUITS WERE BROKEN NEAR THE CHAMBER SIDE DURING USE. IT WAS FURTHER REPORTED THAT FOUR OF THESE FIVE CIRCUITS WERE BROKEN AFTER TWO DISINFECTION CYCLES AND THAT THE TIDAL VOLUME OF THE PATIENT WAS DROPPING WHEN THEY WERE USING THE VENTILATOR PB840 WITH THE CIRCUIT. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT FIVE 900MR810 REUSABLE ADULT BREATHING CIRCUITS WERE BROKEN NEAR THE CHAMBER SIDE DURING USE. IT WAS FURTHER REPORTED THAT FOUR OF THESE FIVE CIRCUITS WERE BROKEN AFTER TWO DISINFECTION CYCLES AND THAT THE TIDAL VOLUME OF THE PATIENT WAS DROPPING WHEN THEY WERE USING THE VENTILATOR PB8400 WITH THE CIRCUIT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 132923 | REUSABLE ADULT BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | 900MR810 | 140404 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | VENTILATOR PB8400| VENTILATOR PB8400 |