FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4542816 · Received February 24, 2015

Report

Report Number
2531779-2015-06493
Event Type
Malfunction
Date Received
February 24, 2015
Report Date
February 9, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 02/09/2015 WITH THE FOLLOWING FINDINGS: THE BATTERY COMPARTMENT WAS CRACKED BELOW THE BUMPER PAD. UNRELATED TO THE BATTERY COMPARTMENT, THE AUDIO BOLUS BUTTON WAS FOUND TO BE DAMAGED, BUT STILL NORMALLY RESPONSIVE TO USER INPUT. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT WAS DAMAGED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 02/09/2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
130301 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 19 YR