FDA Adverse Event Malfunction Summary report: N

CENTRA

MDR report key: 453916 · Received April 8, 2003

Report

Report Number
MW1028123
Event Type
Malfunction
Date Received
April 8, 2003
Date of Event
March 19, 2003
Report Date
April 9, 2003
Manufacturer
HILL-ROM
Product Code
FNL
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

FIRE WITH HILL-ROM 850 ELECTRIC BED. BED FIRE EXPERIENCED ON HILL-ROM MODEL 850 ELECTRIC BED IN 2003. PT'S PARENT HEARD LOUD NOISE FROM BED AND OBSERVED SMOKE AND FLAMES FROM UNDER BED. PT MOVED TO SAFETY WITHOUT HARM, FIRE EXTINGUISHED AND BED REMOVED FROM PT AREA. INSPECTION REVEALED THAT MOTOR CAPACITOR HAD BLOWN AND EITHER IGNITED CAPACITOR CONTENTS OR SHORTED OTHER COMPONENTS. HILL-ROM THEN INFORMED RPTR THAT THEY HAD NOTIFIED CUSTOMERS IN A LETTER DATED JULY 15, 1991 THAT THIS TYPE OF FAILURE WAS POSSIBLE AND THAT THEY "ENCOURAGED" CUSTOMERS TO REPLACE ALL CAPACITORS WITH A NEWER STYLE. RPTR HAS NO RECORD OF RECEIPT OF THAT LETTER AND THERE WAS NEVER ANY ALERT FROM ECRI, FDA, OR OTHER WATCHDOG AGENCIES. REVIEW OF MAUDE DATABASE SHOWED ONE SIMILAR INCIDENT. RPTR HAD OVER 60 BEDS OF THIS MODEL AND FOUND SEVERAL WITH OLD STYLE CAPACITORS. THE MANUFACTURER'S SERVICE REP WAS ALSO NOT AWARE THAT THE CAPACITORS SHOULD HAVE BEEN UPGRADED. SEVERAL OTHER BED MODEL NUMBERS FROM HILL-ROM ARE ALSO AFFECTED (450, 550, 715, 720, 815, 820, 835, 840, 850, 894, 1120, 1130, 1140).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA ELECTRIC BED FNL HILL-ROM 850 *

Patients

Seq Age Sex Outcome Treatment
1 *