TRIO CONNECTOR ADAPTOR
Report
- Report Number
- 0009617544-2015-00076
- Event Type
- Malfunction
- Date Received
- February 23, 2015
- Date of Event
- January 27, 2015
- Report Date
- January 27, 2015
- Manufacturer
- STRYKER SPINE-FRANCE
- Product Code
- MNH
- PMA / PMN Number
- K103292
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
UPDATED LOT CODE: 133835. METHOD: VISUAL INSPECTION; FUNCTIONAL INSPECTION; DEVICE HISTORY REVIEW; COMPLAINT HISTORY REVIEW; RISK ASSESSMENT; RESULTS: MATERIAL ANALYSIS WAS PERFORMED IT WAS DETERMINED THAT GALLING WAS THE CAUSE OF THE SEIZURE OF THE LOCKING NUT. ADDITIONALLY, NO MANUFACTURING DEFECTS WERE OBSERVED DURING THE ANALYSIS. REMOVAL OF THE SEIZED LOCKING NUT LIKELY CAUSED THE DEFORMATION OBSERVED ON THE THREADS. CONCLUSION: THE MOST LIKELY CAUSE OF THE REPORTED EVENT IS GALLING BETWEEN THE SCREW AND NUT THREADS CAUSING THE DEVICE TO JAM.
IT WAS REPORTED THAT; DURING TRIO TRAUMA SURGERY, THE SURGEON CORRECTED KYPHOSIS OF THE PATIENT. WHEN THE SURGEON TRIED TO TIGHTEN THE NUT OF CONNECTOR ADAPTER, THE NUT DID NOT TIGHTEN. THEREFORE, THE SURGEON TRIED TO LOOSEN THE NUT OF CONNECTOR ADAPTER, BUT THE SURGEON COULD NOT LOOSEN. THE SURGEON CHANGED THE CONNECTOR ADAPTER TO ANOTHER CONNECTOR ADAPTER.
IT WAS REPORTED THAT; DURING TRIO TRAUMA SURGERY, THE SURGEON CORRECTED KYPHOSIS OF THE PATIENT. WHEN THE SURGEON TRIED TO TIGHTEN THE NUT OF CONNECTOR ADAPTER, THE NUT DID NOT TIGHTEN. THEREFORE, THE SURGEON TRIED TO LOOSEN THE NUT OF CONNECTOR ADAPTER, BUT THE SURGEON COULD NOT LOOSEN. THE SURGEON CHANGED THE CONNECTOR ADAPTER TO ANOTHER CONNECTOR ADAPTER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 125112 | TRIO CONNECTOR ADAPTOR | PEDICLE SCREW SPINAL SYSTEM | MNH | STRYKER SPINE-FRANCE | 133835 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |