PULSAR 2 GENERATOR
Report
- Report Number
- 1226420-2015-00031
- Event Type
- Malfunction
- Date Received
- February 23, 2015
- Date of Event
- January 26, 2015
- Report Date
- May 29, 2015
- Manufacturer
- MEDTRONIC ADVANCED ENERGY (SALIENT)
- Product Code
- GEI
- PMA / PMN Number
- K102029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PRODUCT EVENT # 700842528 EVALUATION CODE (METHOD): GENERATOR RETURNED AND PENDING ANALYSIS RESULTS EVALUATION CODE (RESULT): GENERATOR RETURNED AND PENDING ANALYSIS RESULTS EVALUATION CODE (CONCLUSION): GENERATOR RETURNED AND PENDING ANALYSIS RESULTS. (B)(4).
(B)(4). EVALUATION PROCESS: *PERFORMED VISUAL INSPECTION ON UNIT PER TCL-514-900008. -UNIT IS IN GOOD CONDITION. *PERFORMED BASELINE FUNCTIONAL TESTING PER TCL-514-900008. -NO ISSUES FOUND. UNIT DELIVERED CORRECT LEVELS OF RF ENERGY IN ALL MODES. UNIT WAS CONFIGURED AS DESCRIBED IN MEDTRONIC DOCUMENT #70-10-1231 (¿PULSAR GENERATOR PREVENTATIVE MAINTENANCE GUIDE¿) FOR MEASURING PEAK CUT 5 ENERGY. THE GENERATOR WAS FOUND TO DELIVER 23.9W OF ENERGY, WITHIN THE TOLERANCE BAND OF +/- 20% FROM THE TARGET POWER OF 20W. ROOT CAUSE: SERVICE DEPARTMENT COULD NOT DUPLICATE ISSUE. UNIT DELIVERED CORRECT LEVELS OF RF ENERGY IN ALL MODES, INCLUDING CUT 5 WITH A 3.0S HANDPIECE. (B)(4).
CUSTOMER WAS PERFORMING INCOMING INSPECTION OF PULSAR GENERATOR AND PLASMABLADE DEVICE AND NOTED THE GENERATOR HAD HIGH OUTPUT ON CUT SETTING 5. THE GENERATOR SHOULD BE GIVING 20 WATTS WITH AND IMPEDANCE OF 100 OHMS, +/- 20%, HOWEVER THE GENERATOR WAS GIVING 86.3 WATTS AT 100 OHMS. CUSTOMER WAS TESTING WITH 3.0S AND FLUKE RF303 ESU ANALYZER. NO PATIENT INVOLVEMENT.
CUSTOMER WAS PERFORMING INCOMING INSPECTION OF PULSAR GENERATOR AND PLASMABLADE DEVICE AND NOTED THE GENERATOR HAD HIGH OUTPUT ON CUT SETTING 5. THE GENERATOR SHOULD BE GIVING 20 WATTS WITH AND IMPEDANCE OF 100 OHMS, +/- 20%, HOWEVER THE GENERATOR WAS GIVING 86.3 WATTS AT 100 OHMS. CUSTOMER WAS TESTING WITH 3.0S AND FLUKE RF303 ESU ANALYZER. NO PATIENT INVOLVEMENT.