ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2015-00090
- Event Type
- Malfunction
- Date Received
- February 19, 2015
- Date of Event
- January 16, 2015
- Report Date
- January 22, 2015
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT212 ADULT INSPIRATORY-HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: ONE OF THE COMPLAINT RT212 BREATHING CIRCUITS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED, PRESSURE TESTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST REVEALED THAT THE BREATHING CIRCUIT WAS OUT OF SPECIFICATION DUE TO EXCESSIVE LEAK. WHEN THE SUBJECT BREATHING CIRCUIT WAS IMMERSED IN THE WATER BATH, THE LEAK WAS OBSERVED AT THE CONNECTION BETWEEN THE LID AND THE BOWL OF THE EXPIRATORY WATER TRAP. A LOT CHECK REVEALED FOUR OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130405. CONCLUSION: THE BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND A LID, WHICH CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED AFTER THE CIRCUIT WAS RELEASED FOR DISTRIBUTION MOST LIKELY DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT212 ADULT INSPIRATORY HEATED BREATHING CIRCUIT STATE THE FOLLOWING: - CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE. - PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. - SET APPROPRIATE VENTILATOR ALARMS. THE HOSPITAL STAFF CORRECTLY CHECKED THE RT212 BREATHING CIRCUIT BEFORE PATIENT USE, WHICH IS IN LINE WITH OUR USER INSTRUCTIONS.
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT TWO RT212 ADULT BREATHING CIRCUITS FAILED THE SERVO-I VENTILATOR LEAK TEST. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 118507 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT212 | 130405 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |