FDA Adverse Event Malfunction Summary report: N

ANSPACH EMAX 2 PLUS BURR MOTOR

MDR report key: 4528939 · Received February 19, 2015

Report

Report Number
3005985723-2015-00018
Event Type
Malfunction
Date Received
February 19, 2015
Date of Event
December 22, 2014
Report Date
January 20, 2015
Manufacturer
MAKO SURGICAL CORP.
Product Code
OLO
PMA / PMN Number
K112507
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPORTED EVENT: ANSPACH EMAX 2 PLUS BURR MOTOR SMOKES AND HAS LOSS OF POWER. THE EVENT WAS CONFIRMED. METHOD & RESULTS: -DEVICE EVALUATION AND RESULTS: DEVICE EVALUATION WAS PERFORMED AND THE ANSPACH EMAX 2 PLUS BURR MOTOR EVENT WAS CONFIRMED. -DEVICE HISTORY REVIEW: NOT PERFORMED AS THE ANSPACH EMAX 2 PLUS BURR MOTOR IS AN OEM PRODUCT. -COMPLAINT HISTORY REVIEW: BASED ON THE DEVICE IDENTIFICATION, THE CATSWEB AND TRACKWISE COMPLAINT DATABASES WERE REVIEWED FROM 2011 TO PRESENT FOR SIMILAR REPORTED EVENTS REGARDING OVERHEATING FAILURE OF P/N 110940 S/N: (B)(4). THERE HAVE BEEN NO OTHER EVENTS FOR THE REFERENCED SERIAL NUMBER. TREND REQUEST FOR THIS PART NUMBER HAS BEEN SUBMITTED (TREND REQUEST #737). CONCLUSIONS: THE ANSPACH EMAX 2 PLUS BURR MOTOR IS AN OEM DEVICE. UPON RECEIPT, THE DEVICE WAS EVALUATED PER TSP-0011 ANSPACH EMAX 2 PLUS BURR MOTOR AND THE REPORTED EVENT WAS CONFIRMED.

Additional Manufacturer Narrative · 1

AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED AT MAKO SURGICAL. EVALUATION OF THE RETURNED MOTORS IS ONGOING. A SUPPLEMENTAL REPORT WILL BE FILED IF ADDITIONAL RESULTS ARE OBTAINED.

Description of Event or Problem · 1

THE SURGEON WAS PERFORMING A BILATERAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE BURR MOTOR BEGAN SMOKING AND LOST POWER.

Description of Event or Problem · 1

THE SURGEON WAS PERFORMING A BILATERAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE BURR MOTOR BEGAN SMOKING AND LOST POWER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
119462 ANSPACH EMAX 2 PLUS BURR MOTOR STEROTAXIC DEVICE, ACCESSORY OLO MAKO SURGICAL CORP.

Patients

Seq Age Sex Outcome Treatment
1 Other