JOHN BUNN NEB-U-LITE EV
Report
- Report Number
- 2428983-2014-00001
- Event Type
- Malfunction
- Date Received
- February 4, 2015
- Date of Event
- May 7, 2014
- Report Date
- February 4, 2015
- Manufacturer
- ZONGSHAN A&J MEDICAL EQUIPMENT CO., LTD.
- Product Code
- CAF
- Product Problem
- Yes
- Report Source
- Distributor report
- Reporter Occupation
- OTHER
Narratives
THE MANUFACTURER SUBMITTED THE RESULTS OF THEIR INVESTIGATION OF THE ACTUAL RETURNED UNIT IN ADDITION TO THE RESULTS OF AN EXEMPLAR UNIT USED TO DUPLICATE THE FAILURE. CONCLUSION: FAILURE DUE TO INCORRECT POWER SUPPLY/VOLTAGE. THE MANUFACTURER RETURNED THE ABOVE TESTED UNITS TO GF HEALTH PRODUCTS, INC. (GFHP). ON 01/09/2015 GFHP DELIVERED BOTH UNITS TO APPLIED TECHNICAL SERVICES, INC. (THIRD PARTY DOMESTIC TESTING LAB) FOR VALIDATION AND FURTHER ANALYSIS. ATS TEST REPORT CONCLUSION: FAILURE OF CUSTOMER DAMAGED UNIT WAS MOST LIKELY DUE TO OPERATING VOLTAGE BEING TOO HIGH, WHICH CAUSED A BREAKDOWN THROUGH THE VARNISH OF THE WINDINGS. THE DAMAGED MOTOR WAS NOT DUE TO A MANUFACTURING DEFECT. ROOT CAUSE OF FAILURE: END USER/CUSTOMER MISUSE. THE UNIT WAS PLUGGED INTO A 220 POWER VOLTAGE. THE LABEL ON THE UNIT AND THE USER MANUAL SPECIFIES 120V. CONCLUSIONS: DAMAGE TO THE MOTOR, INCLUDING BURN MARKS ON WINDINGS AND WARPING OF WINDING HOUSING TO BOTH UNITS WERE ALMOST IDENTICAL. MULTIPLE BURN MARKS ON CUSTOMER DAMAGED MOTOR INDICATE FAILURE WAS NOT CAUSED BY A SINGLE-POINT BREAKDOWN IN THE VARNISH DUE TO MANUFACTURER DEFECT. THERMAL PROTECTION SWITCH IN BOTH UNITS WAS STILL INTACT. MOTORS DID NOT GET HOT ENOUGH FOR VARNISH TO FAIL SIMPLY DUE TO OVERHEATING. FAILURE OF CUSTOMER DAMAGED UNIT WAS MOST LIKELY DUE TO OPERATING VOLTAGE BEING TOO HIGH, WHICH CAUSED A BREAKDOWN THROUGH THE VARNISH OF THE WINDINGS.
DETAILED IN 30-DAY INITIAL REPORT TO THE FDA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 83495 | JOHN BUNN NEB-U-LITE EV | NEBULIZER COMPRESSER | CAF | ZONGSHAN A&J MEDICAL EQUIPMENT CO., LTD. | JB0112-061 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |