S5 DOUBLE HEAD PUMP
Report
- Report Number
- 9611109-2015-00017
- Event Type
- Malfunction
- Date Received
- February 13, 2015
- Date of Event
- January 14, 2015
- Report Date
- March 8, 2017
- Manufacturer
- SORIN GROUP DEUTSCHLAND
- Product Code
- DTQ
- PMA / PMN Number
- K071318
- Removal / Correction Number
- Z-0956/0965-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- UNKNOWN
Narratives
PATIENT INFO WAS NOT PROVIDED. SORIN GROUP (B)(4) MFRS THE S5 DOUBLE HEAD PUMP. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP RECEIVED A REPORT THAT THE TOUCH SCREEN OF THE S5 DOUBLE HEAD PUMP WAS UNRESPONSIVE DURING A PREVENTATIVE MAINTENANCE BY THE SGU FIELD SERVICE REP. THERE WAS NO PATIENT INVOLVEMENT. THE INVESTIGATION IS ONGOING. A F/U REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.
RECALL: THERE HAVE BEEN TWO FIELD ACTIONS FOR THIS ISSUE. THE Z-NUMBERS ARE Z-0956/0965-2011 AND Z-1149/1158-2012. LIVANOVA (B)(4) MANUFACTURES THE S5 DOUBLE HEAD PUMP. THE EVENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE. THE SERVICE REPRESENTATIVE WAS ABLE TO REPRODUCED THE REPORTED ISSUE AND REPLACED THE FAULTY LCD TOUCH SCREEN. THE UNIT WAS TESTED WITHOUT ISSUE AND THE SOFTWARE WAS UPGRADED. PREVENTATIVE MAINTENANCE AND A FUNCTIONAL VERIFICATION WERE PERFORMED AND THE UNIT WAS RETURNED TO SERVICE. THE REPLACED TOUCH SCREEN WAS RETURNED TO SORIN GROUP USA FOR FURTHER INVESTIGATION, WHERE A PHOTOGRAPH OF THE KAPTON-TAIL DATE-CODE WAS TAKEN. THE PHOTO WAS PROVIDED TO LIVANOVA (B)(4) FOR REVIEW. ANALYSIS OF THE PHOTOGRAPH BY LIVANOVA (B)(4) CONFIRMED AGING AND WEAR TO BE THE MOST LIKELY ROOT CAUSE. A REVIEW OF THE DHR DID NOT IDENTIFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. LIVANOVA (B)(4) INITIATED A ROOT CAUSE INVESTIGATION ((B)(4)) FOR THIS TYPE OF ISSUE.
SORIN GROUP RECEIVED A REPORT THAT THE TOUCH SCREEN OF THE S5 DOUBLE HEAD PUMP WAS UNRESPONSIVE DURING A PREVENTATIVE MAINTENANCE BY THE SGU FIELD SERVICE REP. THERE WAS NO PATIENT INVOLVEMENT.