ENSEAL G2 STRAIGHT JAW
Report
- Report Number
- 3005075853-2015-01122
- Event Type
- Malfunction
- Date Received
- February 17, 2015
- Report Date
- December 11, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K112033
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ASKED FOR BUT UNKNOWN OR NOT PROVIDED DURING INITIAL CONTACT. SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THE DEVICE WAS RECEIVED WITH THE TOP OF THE I-BLADE FRACTURED AND SLIGHTLY LIFTED. IN ADDITION THE UPPER JAW WAS DETACHED AND RETURNED, ALSO THE ENERGY BUTTON DETACHED AND RETURNED WITH THE INSTRUMENT. THE BUTTON WAS REATTACHED TO THE INSTRUMENT AND CONNECTED TO THE GEN11 AND IT WAS RECOGNIZED. BECAUSE OF THE CONDITION OF THE DEVICE NOT ALL FUNCTIONAL TESTING COULD BE PERFORMED WITH THE GENERATOR. THE CONDITION OF THE BLADE PREVENTED THE FUNCTIONALITY OF THE JAW. THE JAW COULD NOT BE HELD IN PLACE DUE TO THE BREAKAGE. A PROBABLE CAUSE OF THE DAMAGE TO THE I BLADE COULD BE NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT. ALTHOUGH NO CONCLUSION COULD BE REACHED AS TO HOW THE BUTTON DETACHED FROM THE INSTRUMENT IT SHOULD BE NOTED THAT OUR MANUFACTURING PROCESS VERIFIES THE PRESENCE AND FUNCTIONALITY OF THE INSTRUMENT PRIOR TO SHIPPING.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SIGMA PROCEDURE, THE JAW BROKEN AFTER A FEW TIMES; PART FELL INTO SITUS BUT COULD BE REMOVED. NO FURTHER INFORMATION IS AVAILABLE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 114159 | ENSEAL G2 STRAIGHT JAW | ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | NA | L4EG4X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |