JETSTREAM XC-2.4 CATHETER
Report
- Report Number
- 2183460-2015-00012
- Event Type
- Malfunction
- Date Received
- February 10, 2015
- Date of Event
- January 16, 2015
- Report Date
- January 19, 2015
- Manufacturer
- BOSTON SCIENTIFIC
- Product Code
- MCW
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
UDI - (B)(4). QUALITY ASSURANCE PRODUCT ANALYSIS RECEIVED AND EXAMINED THE RETURNED JETSTREAM XC-2.4 CATHETER, LOT NUMBER 183343. VISUAL EXAMINATION DETERMINED THAT THE MAILMAN GUIDEWIRE WAS STUCK INSIDE THE CATHETER. THE REMAINDER OF THE DEVICE APPEARED TO BE IN GOOD OVERALL CONDITION. THE RETURNED UNIT WAS FUNCTIONALLY TESTED AND PASSED ALL REQUIREMENTS. FURTHER EXAMINATION OF THE GUIDEWIRE LUMEN CONFIRMED THAT THE BUSHING WAS OCCLUDED WITH TEFLON AND WAS NOT ABLE TO BE REMOVED. PRODUCT ANALYSIS CONCLUDED THAT THE GUIDEWIRE LOST COATING AND OCCLUDED THE BUSING AT THE DISTAL END OF THE CATHETER THEREBY CONTRIBUTING TO THE LOSS OF TRACTABILITY OF THE GUIDEWIRE.
THE SITE REPORTED THE FOLLOWING: THE JETSTREAM ATHERECTOMY CATHETER WAS BEING USED TO DEBULK AN OCCLUDED VESSEL. AFTER SEVERAL PASSES, THE CATHETER WAS REMOVED AND IMAGES WERE OBTAINED. THE PHYSICIAN DECIDED THAT HE WANTED TO PERFORM ATHERECTOMY ONCE AGAIN. HOWEVER, THE DEVICE BECAME STUCK ON THE MAILMAN GUIDEWIRE AND WAS NOT ABLE TO BE ADVANCED. THE CATHETER WAS REMOVED AND THE PHYSICIAN ATTEMPTED TO USE A SECOND JETSTREAM CATHETER BUT IT BECAME STUCK ON THE GUIDEWIRE AS WELL. THE PHYSICIAN REMOVED THE JETSTREAM CATHETER WITH THE GUIDEWIRE, DECIDED THAT THE TREATMENT WAS SUCCESSFUL AND COMPLETED THE PROCEDURE. NO ADVERSE EVENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 97276 | JETSTREAM XC-2.4 CATHETER | ATHERECTOMY CATHETER | MCW | BOSTON SCIENTIFIC | 112266-001 | 183343 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |