FDA Adverse Event
Malfunction
Summary report: N
TDXSP ELECTRIC RED EURO WIDE BASE W/ 23-1 SSD GEAR BOX
MDR report key: 4509940
·
Received February 12, 2015
Report
- Report Number
- 1031452-2015-07133
- Event Type
- Malfunction
- Date Received
- February 12, 2015
- Date of Event
- January 30, 2015
- Report Date
- February 5, 2015
- Manufacturer
- INVACARE FLORIDA OPERATIONS
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Description of Event or Problem · 1
CHAIR TDXSP WHEN POWERING UP DURING NEUTRAL TESTING MODE IF JOYSTICK IS TOUCHED CHAIR TAKES OFF . CHAIR HAS ASL MONSTER JOYSTICK AND SIDE EGG SWITCHES THAT WERE ATTACHED TO HOUSING AFTER DELIVERY WARRANTY SEAL WAS BROKEN ON JSK. NO INJURY OR DAMAGE USER GOT STARTLED. THIS EVENT HAPPENED ONCE FOR THE USER AND ONCE FOR THE DEALER WITHIN A DAY OUR DEALER HAS CHAIR AND CAN NOT DUPLICATE THE PROBLEM AFTER MULTIPLE HOURS OF TESTING
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 103890 | TDXSP ELECTRIC RED EURO WIDE BASE W/ 23-1 SSD GEAR BOX | WHEELCHAIR, POWERED | ITI | INVACARE FLORIDA OPERATIONS | TDXSP-CG |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |