PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4509472
·
Received February 11, 2015
Report
- Report Number
- 2032227-2015-06286
- Event Type
- Malfunction
- Date Received
- February 11, 2015
- Date of Event
- January 28, 2015
- Report Date
- January 28, 2015
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH HIGH IDLE CURRENT DUE TO MOISTURE DAMAGE FOUND ON INTERFACE BOARD. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW AND CRACKED RESERVOIR TUBE LIP. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
Description of Event or Problem · 1
IT WAS REPORTED ON (B)(6) 2014, THE INSULIN PUMP HAD A LOW BATTERY LIFE. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. NO FURTHER INFORMATION REPORTED.