FDA Adverse Event Malfunction Summary report: N

TROCAR SLEEVE 10/110MM SMOOTH WITH TAP

MDR report key: 4507189 · Received February 3, 2015

Report

Report Number
2916714-2015-00061
Event Type
Malfunction
Date Received
February 3, 2015
Report Date
February 3, 2015
Manufacturer
AESCULAP AG & CO. KG
Product Code
GCJ
PMA / PMN Number
K101937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

U.S. AGENT NOTIFIED ON: (B)(4) 2015. MANUFACTURING SITE EVALUATION: INSTRUMENTS HAVE BEEN CHECKED MICROSCOPICALLY. NO MICRO CRACKS OR OTHER DAMAGES COULD BE FOUND. THE MONOPOLAR INSTRUMENTS WERE CHECKED IN THE MANUFACTURING DEPARTMENT REGARDING ELECTRICAL DISCHARGE. ALL INSTRUMENTS WERE FOUND TO BE WITHOUT FAILURES. NO DEVIATIONS FROM OEM SPECIFICATIONS COULD BE VERIFIED. THERE ARE NO INDICATIONS FOR MATERIAL OR PRODUCT DEVIATIONS.

Description of Event or Problem · 1

COUNTRY OF COMPLAINT: (B)(6). CUSTOMER STATES INSTRUMENTS ARE NOT IN GOOD CONDITION, AND THE SHEATHS MAY HAVE MICRO CRACKS. THERE WAS A LESION, BY BURN, IN THE AREA OF SURGERY. INSTRUMENTS FROM SET INCLUDE: PO004R METZENBAUM SCISSORS 5MM 310MM (PO603R; PM973R; PO958R); PO102R MARYLAND DISSECTOR 5MM 310MM (PO608R; PM973R; PO958R); PO128R CLINCHING GRASPER FENEST 5MM 310MM (PO770R; PM973R; PO959R); PO191R UNIVERSAL GRASPER 5MM 3 (PO736R; PM973R; PO959R); GK373R INNER TUBE W/HANDLE FOR GK372R; GK384R CERAMIC ELECTRODE TIP L-HK F/GK372R; EK024R TROCAR SLEEVE 10/110MM SMOOTH WITH TAP; EK014R TROCAR SLEEVE 5/110MM SMOOTH WITH TAP; EK064R TROCAR PIN BLUNT (HASSON) 10/110MM; EK046R TROCAR PIN CONICAL SHARP 5/110MM. MEDWATCH FILED ON MOST LIKELY COMPONENT TO HAVE BEEN CAUSED THE BURN. ALL OTHER COMPONENTS WERE USED WITHIN THE CASE AND ARE BEING EVALUATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
78055 TROCAR SLEEVE 10/110MM SMOOTH WITH TAP LAPAROSCOPIC TROCAR GCJ AESCULAP AG & CO. KG EK024R

Patients

Seq Age Sex Outcome Treatment
1 Other