FDA Adverse Event Injury Summary report: N

ABBOTT PLUM +

MDR report key: 450421 · Received March 25, 2003

Report

Report Number
450421
Event Type
Injury
Date Received
March 25, 2003
Date of Event
March 5, 2003
Report Date
March 11, 2003
Manufacturer
ABBOTT LABORATORIES, HOSPITAL PRODUCT DIVISION
Product Code
FRN
Adverse Event
Yes
Report Source
User Facility report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

MORPHINE DRIP-50 MG IN 50CC'S-WAS HUNG AT 2200. THE RATE WAS TO BE 5 CC/HR. AT 2300, THE BAG WAS EMPTY. THE EXACT RATE THE PUMP WAS SET AT IS NOT CLEAR AT THIS ITME. THE PT RECEIVED 0.4MG OF NARCAN. VITALS REMAINED STABLE BEFORE AND AFTER-RESP RATE OF 12-14 WITH O2 SATS 98 TO 100% ON O2. NO APPARENT HARM. ADDENDUM: MEMORY FROM PUMP DOWNLOADED SHOWS RATE WAS SET AT 50 CC/HR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABBOTT PLUM + INFUSION PUMP FRN ABBOTT LABORATORIES, HOSPITAL PRODUCT DIVISION PLUM A+ *

Patients

Seq Age Sex Outcome Treatment
1 87 YR Required Intervention