FDA Adverse Event
Other
Summary report: N
SPOT PHOTOTHERAPY WALL MOUNT LIGHT
MDR report key: 449554
·
Received March 17, 2003
Report
- Report Number
- 2018492-2003-00009
- Event Type
- Other
- Date Received
- March 17, 2003
- Date of Event
- January 16, 2003
- Report Date
- March 17, 2003
- Manufacturer
- BURTON MEDICAL PRODUCTS CORP.
- Product Code
- FSY
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
END USER REPORTED THAT A PT WHILE BEING TREATED IN A WARMER WITH 2 SPOT PHOTOTHERAPY LIGHTS 12" - 14" FOR A DURATION OF 1 HOUR AND 15 MINUTES REC'D 2ND DEGREE BURNS ON CHEST. PT WAS TREATED WITH BACITRACIN. PLASTIC SURGEON EVALUATED PT AT TIME OF THE EVENT AND INDICATED NO SCARRING WOULD RESULT. 4 DAYS LATER BURN HEALED WITH NO VISIBLE SCARRING. A BIOMED TECH EVALUATED BOTH LAMPS AND FOUND THAT THEY PASSED ALL SAFETY TESTS: OUTPUT AT 20" WAS FINE, LAMPS FUNCTIONED PERFECTLY, LAMPS WERE IN EXCELLENT WORKING ORDER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPOT PHOTOTHERAPY WALL MOUNT LIGHT | WALL MOUNT | FSY | BURTON MEDICAL PRODUCTS CORP. | 6600-0288-901 | 1-99BURT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Other |