FDA Adverse Event Other Summary report: N

SPOT PHOTOTHERAPY WALL MOUNT LIGHT

MDR report key: 449554 · Received March 17, 2003

Report

Report Number
2018492-2003-00009
Event Type
Other
Date Received
March 17, 2003
Date of Event
January 16, 2003
Report Date
March 17, 2003
Manufacturer
BURTON MEDICAL PRODUCTS CORP.
Product Code
FSY
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

END USER REPORTED THAT A PT WHILE BEING TREATED IN A WARMER WITH 2 SPOT PHOTOTHERAPY LIGHTS 12" - 14" FOR A DURATION OF 1 HOUR AND 15 MINUTES REC'D 2ND DEGREE BURNS ON CHEST. PT WAS TREATED WITH BACITRACIN. PLASTIC SURGEON EVALUATED PT AT TIME OF THE EVENT AND INDICATED NO SCARRING WOULD RESULT. 4 DAYS LATER BURN HEALED WITH NO VISIBLE SCARRING. A BIOMED TECH EVALUATED BOTH LAMPS AND FOUND THAT THEY PASSED ALL SAFETY TESTS: OUTPUT AT 20" WAS FINE, LAMPS FUNCTIONED PERFECTLY, LAMPS WERE IN EXCELLENT WORKING ORDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPOT PHOTOTHERAPY WALL MOUNT LIGHT WALL MOUNT FSY BURTON MEDICAL PRODUCTS CORP. 6600-0288-901 1-99BURT

Patients

Seq Age Sex Outcome Treatment
1 48 YR Other