FDA Adverse Event Summary report: N

STRYKER INSTRUMENTS

MDR report key: 449247 · Received March 18, 2003

Report

Report Number
MW1027929
Date Received
March 18, 2003
Date of Event
March 17, 2003
Report Date
March 18, 2003
Manufacturer
STRYKER INSTRUMENTS
Product Code
GEI
Report Source
Voluntary report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

ADD'L INFO RECEIVED FROM MFR 5/22/03: TEST SUMMARY READS: THE TEST WAS COMPLETED AS SPECIFIED IN THE TEST PROCEDURE WITH BOTH BURS MEETING THE MINIMUM PERFORMANCE REQUIRED. MANUAL TESTING WAS PERFORMED AS DETAILED ABOVE, THE BURS DID NOT EXHIBIT EXCESSIVE HEAD DETERIORATION, SHELL REMOVAL OF COATING, EXCESSIVE LOSS OF DIAMOND COATING OR FATIGUE. THERE WERE NO BREAKAGES. THERE WAS NOTICEABLE DISCOLORATION HOWEVER THIS MAY BE DUE TO THE EXTENDED PERIOD CUTTING WITHOUT IRRIGATION.

Description of Event or Problem · 1

DIAMOND COATING CAME OFF DRILL BIT. CAME OFF IN PT BUT WAS ABLE TO BE RETRIEVED BY SURGEON. NO ADVERSE EFFECT TO PT. SAME PROBLEM OCCURRED TWICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STRYKER INSTRUMENTS DIAMOND COATED DRILL BIT GEI STRYKER INSTRUMENTS * 021024478M

Patients

Seq Age Sex Outcome Treatment
1 *