I-STAT ACT KAOLIN CARTRIDGE
Report
- Report Number
- 2245578-2015-00005
- Event Type
- Malfunction
- Date Received
- February 6, 2015
- Date of Event
- January 13, 2015
- Report Date
- January 14, 2015
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JBP
- PMA / PMN Number
- K023582
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4).
(B)(4). THE INVESTIGATION WAS COMPLETED ON 04/14/2015. RETAIN AND RETURNED PRODUCT WAS TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
NA
ON (B)(6) 2015, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGE THAT YIELDED UNEXPECTED RESULTS ON A PATIENT UNDERGOING A PROCEDURE IN THE CATH LAB. CUSTOMER IS RETURNING PRODUCT FOR INVESTIGATION. TIME I-STAT I-STAT HEPARIN HEPARIN TIME0754 132 10,000 PRIME UNK0836 337 25,000 08330849 299 10,000 08450916 420 376 10,000 09100927 321 - - -0938 426 5,000 09351000 243 243 - -1015 >1000 10,000 10071045 376 >1000 - -1110 359 315 - -1125 - - 5,000 11201127 382 276 - -1145 277 282 5,000 11401200 >1000 243 - -1225 293 >1000 7,000 12081250 265 288 - -1315 193 - - -THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 87889 | I-STAT ACT KAOLIN CARTRIDGE | ACT KAOLIN CARTRIDGE | JBP | ABBOTT POINT OF CARE | NA | T14259A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |