FDA Adverse Event Malfunction Summary report: N

I-STAT ACT KAOLIN CARTRIDGE

MDR report key: 4491798 · Received February 6, 2015

Report

Report Number
2245578-2015-00005
Event Type
Malfunction
Date Received
February 6, 2015
Date of Event
January 13, 2015
Report Date
January 14, 2015
Manufacturer
ABBOTT POINT OF CARE
Product Code
JBP
PMA / PMN Number
K023582
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 04/14/2015. RETAIN AND RETURNED PRODUCT WAS TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Description of Event or Problem · 1

NA

Description of Event or Problem · 1

ON (B)(6) 2015, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING ACT KAOLIN CARTRIDGE THAT YIELDED UNEXPECTED RESULTS ON A PATIENT UNDERGOING A PROCEDURE IN THE CATH LAB. CUSTOMER IS RETURNING PRODUCT FOR INVESTIGATION. TIME I-STAT I-STAT HEPARIN HEPARIN TIME0754 132 10,000 PRIME UNK0836 337 25,000 08330849 299 10,000 08450916 420 376 10,000 09100927 321 - - -0938 426 5,000 09351000 243 243 - -1015 >1000 10,000 10071045 376 >1000 - -1110 359 315 - -1125 - - 5,000 11201127 382 276 - -1145 277 282 5,000 11401200 >1000 243 - -1225 293 >1000 7,000 12081250 265 288 - -1315 193 - - -THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
87889 I-STAT ACT KAOLIN CARTRIDGE ACT KAOLIN CARTRIDGE JBP ABBOTT POINT OF CARE NA T14259A

Patients

Seq Age Sex Outcome Treatment
1 67 YR