FDA Adverse Event Malfunction Summary report: N

ONE TOUCH BASIC

MDR report key: 449109 · Received March 17, 2003

Report

Report Number
2939301-2003-02085
Event Type
Malfunction
Date Received
March 17, 2003
Report Date
March 12, 2003
Manufacturer
LIFESCAN, INC.
Product Code
CFR
Removal / Correction Number
NA
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

A PATIENT REPORTED BLOOD GLUCOSE RESULTS OF 212, 152 MG/DL AND 153 MG/DL WITH A LIFESCAN METER, PEFORMED WITHIN 10 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS THE EXPECTED VALUE OF <=20% AND/OR <=20 MG/DL, THEREBY INDICATING A POTENTIAL MALFUNCTION OF THE METER IN TERMS OF PRECISION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH BASIC BLOOD GLUCOSE MONITORING KIT/SYSTEM CFR LIFESCAN, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 60 YR